3.5 Installing The Defi btech Data Card (DDC Card)
(optional)
The Defi btech Data Card (DDC card) is used to store event and audio information collected by the AED. All
DDU-2000 Series AEDs will operate without DDC cards and will still store select event information internally.
Information stored on the DDC card is retrievable with a separate Defi btech PC-based software package. (Refer to
the “
Defi bView
” section in Chapter 8 of this manual.)
Before installing the DDC card, ensure the AED is turned OFF. Locate the data card/USB port access door on the
right-hand side of the unit. Open the data card/USB port access door by slightly pushing and then sliding the door
down to release the latch. The door will spring open. Insert the DDC card into the thin slot in the side of the AED
centered above the USB port opening, notched end fi rst, label side up, until it clicks into place. The card should be
fl ush with the surface of the slot. If the card does not push in all the way, it may have been inserted upside down.
In that case, remove the card, fl ip it over, and try inserting it again.
To remove the DDC card, press the card as far as it will go and then release. Upon release, the DDC card will be
partially ejected and can be removed by pulling the DDC card the rest of the way out.
Close the data card/USB port access door by closing and then pushing the door up until the door latch engages.
Note
: DDU-2000 Series AEDs running software version 1.3 or below may not function with certain DDC Cards.
Contact Defi btech for support (refer to Chapter 12 for contact information).
CAUTION
Using non-Defi btech Data Cards
(DDC cards) may damage the unit
and will void the warranty.
3.6 Completing The Installation
Once you have completed the previous steps to set up your DDU-2000 Series AED, follow this procedure:
1. Turn the unit on by pressing the ON/OFF button.
2. Listen for the “
Call for Help
” voice prompt.
3. Turn the unit off by pressing and holding the ON/OFF button.
4. Listen for the “
Powering Off
” voice prompt.
5. Check Active Status Indicator (ASI) to verify that it is fl ashing green.
(Refer to the ”
Self-Tests”
section in Chapter 5 of this manual for instructions on how to run a manually initiated self-test.)
3.7 Storing The DDU-2000 Series AED
Store the DDU-2000 Series AED, with defi brillation pads attached, in environmental conditions within range of the
specifi cations. (Refer to the “
Environmental
” section in Chapter 9 of this manual.) The unit should also be stored so
that the Active Status Indicator (ASI) can be readily seen.
The Active Status Indicator (ASI) should periodically fl ash with a green light. If it fl ashes with a red light or does not
fl ash at all, the DDU-2000 Series AED needs servicing. (Refer to Section 3.4, “
Checking The DDU-2000 Series AED
Status
”, for more information.)
Defi btech recommends storing your AED in an easily accessible location where the unit can be seen and heard.