Battery Charger Information for FROG

-

4000

TM

 

Battery Pack Type

 

NiMH 

 

Battery Charger Input

 

100

240V AC

 

Battery Pack Voltage

 

9.6V

 

Battery Pack Capacity

 

5000mAh

 

Battery Charger Current Selection (switch)

 

1.8A

 

Charging Time

 

Approximately 210 Minutes

 

FROG

-

4000

TM

 Battery  and Battery Charger Specifications

 

LED Indications

 

 

Green LED

 

Red LED

 

On, Not Flashing

 

Battery is fully charged.

 

Battery is charging.

 

The green light and the red light will flash twice alternately when the charger is powered on.

 

Off

 

The charger is off.  Causes: Short circuit, reversed polarity, battery temperature above 60

o

C, or 4 hours 

safety timer exceeded.  

 

Precautions and Safety Features:

 

· 

Do not use the charger for battery packs with capacity less than 900mAh or greater than 5000mAh.

 

· 

Do not use the charger for battery packs under 9.6V or over 18V.  The charger will shut off if voltage is below 6V.

 

· 

Use charger only for NiMH or NiCd battery packs.  Highly recommend using Tenergy brand, high drain rate battery packs.  Neither 

Defiant Technologies nor Tenergy Corporation are responsible for  damage caused by charging other battery pack brands with this 

charger.

 

· 

Use 1.8A current for battery packs with capacities between 1800mAh and 5000mAh.

 

· 

Limit the duration of charging when using 1.8A current.  

 

· 

The charger automatically shuts off after 4 hours charging.

 

· 

The charger is designed for indoor use.

 

· 

Use the charger in a well

-

ventilated, dry area away from flammable, explosive goods. 

 

· 

The charger is designed to be used on a horizontal surface.

 

Instructions for Charging the Battery:

 

 

 

1.

 

Remove the sparge bottle from the instrument before charging the battery.

 

2.

 

The Instrument should be powered off and disconnected from the power cable.

 

3.

 

Plug battery charger into Battery charge port (S) and outlet source.

 

4.

 

Place battery charger switch to 1.8A.

 

5.

 

While the battery is charging, the light on the charger switch will appear red.

 

6.

 

When the battery is completely charged, the light on the charger switch will appear green.

 

7.

 

It is recommended that the battery be fully charged before the instrument is used in an 

outdoor setting.

 

S, Battery 

To S, 

 

Battery 

Charge

 

To Outlet

 

Содержание FROG-4000

Страница 1: ...wer 4 Turn the instrument power switch to the ON position fully up The instrument display screen will be visible 4 Open Ellvin by double clicking on the Ellvin icon 5 Ellvin opens in the default Live...

Страница 2: ...position Running a Sample 1 Instrument screen reads READY 2 Record Log and other relevant sample data for future reference 3 Press the start button in Ellvin or on the instrument to begin an analysis...

Страница 3: ...files Open the folder and the screen displays a list of sequential files named LOG_XXX 3 Select the LOG file of interest and drag copy it to the computer 4 Open Ellvin 5 Open Live Data Window 6 Click...

Страница 4: ...Look In to select the FOLDER labeled with the DATE that Ellvin created converted the file of interest 4 Click on the file of interest and a preview chromatogram appears on the right side of the scree...

Страница 5: ...t is CONNECTED and whether or not parameters have been updated GC Column Temperature Indicator Displays the GC temperature The icon is GREEN when instrument is ready for loading The icon is RED when t...

Страница 6: ...of the features in Live Data Window the following features are available in Analyze Window Integrator 1 The user creates a baseline and integrates peaks then calculates the peak height and area to det...

Страница 7: ...he analyte in the Analyte Alias window maximum of 5 Characters 3 Click the ADD button Correlation Chart and Equations This feature in the Calibration Window displays the degree of correlation between...

Страница 8: ...s Tb Ramp time from GC cold to GC hot temperature seconds Tc Hold time at hot GC temperature seconds Ct Initial cold GC temperature o C Ht Final hot GC temperature o C CLEAN Time cleaning PC by heatin...

Страница 9: ...air sampler by connecting to this port Figure 5 Connect exhaust connector to exhaust port ANALYZE Position Figure 3 Sample Introduction Valve Coupler Line Transfer Line Sparge Needle Port Valve Contr...

Страница 10: ...ampler can now be connected to a Tedlar bag to calibrate the instrument or to the wand to collect air samples 5 Press the yellow start button on the instrument or in Ellvin to begin an air sample anal...

Страница 11: ...position is facing toward you 5 Turn valve position counterclockwise to the right place sparge bottle in up position on Frog 4000 and tighten the nut around sparge bottle 6 Run Frog 4000 as normal by...

Страница 12: ...d high drain rate battery packs Neither Defiant Technologies nor Tenergy Corporation are responsible for damage caused by charging other battery pack brands with this charger Use 1 8A current for batt...

Страница 13: ...Sparge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter 8 Power Cable 9 Battery Charg...

Страница 14: ...rge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter d Air Sample Wand 8 Power Cable 9...

Отзывы: