FROG

-

4000

TM

 

Soil Sampling 

 

www.defiant

-

tech.com

 

Soil sampling requires a scale, scapula, disposable 

weigh boats. (Figure below)

 

 

1.

 

Measure 1g of soil to be tested in a clean, disposable weigh boat or in a soil syringe.  (Figure 2)

 

2.

 

Add soil directly to sparge bottle (Figure 3)

 

3.

 

Place sparge bottle in down position on Frog

-

4000 

 

4.

 

Add 5mL of DI water using a glass or disposable syringe (Figure 4). Be sure the valve position is facing toward you

 

5.

 

Turn valve position counterclockwise to the right, place sparge bottle in up position on Frog

-

4000, and tighten the nut around 

sparge bottle

 

6.

 

Run Frog

-

4000 as normal by either using “Play” button in Ellvin software, or pressing the yellow button on the side of  Frog

-

4000

 

NOTE: A separate application note on Methanol Extraction can be found on our website www.defiant

-

tech.com/downloads.php

 

Figure 2: 1g of soil to be tested

 

Figure 3: Soil put directly  into sparge bottle

 

Figure 4: Adding 5mL of DI water to soil

 

or

 

Note: 

For Methanol Extraction, 

please visit www.defiant

-

tech.com 

for application note.

 

Содержание FROG-4000

Страница 1: ...wer 4 Turn the instrument power switch to the ON position fully up The instrument display screen will be visible 4 Open Ellvin by double clicking on the Ellvin icon 5 Ellvin opens in the default Live...

Страница 2: ...position Running a Sample 1 Instrument screen reads READY 2 Record Log and other relevant sample data for future reference 3 Press the start button in Ellvin or on the instrument to begin an analysis...

Страница 3: ...files Open the folder and the screen displays a list of sequential files named LOG_XXX 3 Select the LOG file of interest and drag copy it to the computer 4 Open Ellvin 5 Open Live Data Window 6 Click...

Страница 4: ...Look In to select the FOLDER labeled with the DATE that Ellvin created converted the file of interest 4 Click on the file of interest and a preview chromatogram appears on the right side of the scree...

Страница 5: ...t is CONNECTED and whether or not parameters have been updated GC Column Temperature Indicator Displays the GC temperature The icon is GREEN when instrument is ready for loading The icon is RED when t...

Страница 6: ...of the features in Live Data Window the following features are available in Analyze Window Integrator 1 The user creates a baseline and integrates peaks then calculates the peak height and area to det...

Страница 7: ...he analyte in the Analyte Alias window maximum of 5 Characters 3 Click the ADD button Correlation Chart and Equations This feature in the Calibration Window displays the degree of correlation between...

Страница 8: ...s Tb Ramp time from GC cold to GC hot temperature seconds Tc Hold time at hot GC temperature seconds Ct Initial cold GC temperature o C Ht Final hot GC temperature o C CLEAN Time cleaning PC by heatin...

Страница 9: ...air sampler by connecting to this port Figure 5 Connect exhaust connector to exhaust port ANALYZE Position Figure 3 Sample Introduction Valve Coupler Line Transfer Line Sparge Needle Port Valve Contr...

Страница 10: ...ampler can now be connected to a Tedlar bag to calibrate the instrument or to the wand to collect air samples 5 Press the yellow start button on the instrument or in Ellvin to begin an air sample anal...

Страница 11: ...position is facing toward you 5 Turn valve position counterclockwise to the right place sparge bottle in up position on Frog 4000 and tighten the nut around sparge bottle 6 Run Frog 4000 as normal by...

Страница 12: ...d high drain rate battery packs Neither Defiant Technologies nor Tenergy Corporation are responsible for damage caused by charging other battery pack brands with this charger Use 1 8A current for batt...

Страница 13: ...Sparge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter 8 Power Cable 9 Battery Charg...

Страница 14: ...rge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter d Air Sample Wand 8 Power Cable 9...

Отзывы: