6
0
A good reading, but the predicted cured thickness is
less than the minimum range.
+++
A good reading, but the predicted cured thickness is
greater than the maximum range.
---
An indeterminate or invalid reading.
measuring a different location on the part. If a valid measurement
cannot be calculated, one of the following messages will be
displayed:
Single Reading Mode (default)
The gage will automatically display the predicted cured thickness
after analyzing the powder. To take another reading, release the
button, then press and hold the again.
Scan Mode
Press and hold the button. The gage continuously analyzes
the powder and displays the predicted cured thickness until the
button is released. The last valid measurement is added to
memory when the button is released if Memory mode is on.
Maintenance
The probe screen protecting the sensor of the
PosiTest PC
can
be unscrewed for cleaning. Accumulated powder should be
carefully removed from the screen by no more than 0.2 MPa (20 -
30 psi) of compressed air.
- Do not blow air directly into the sensor.
- Replacement screens are available for purchase.