4
For best accuracy, perform a
Zero
at the
beginning of every shift to compensate for
temperature and humidity changes. It is also
required after a Gage Reset or whenever the
“Zero Probe” graphic is displayed.
1.
Select
Zero
from the Gage menu.
2.
Place the probe tip in the supplied
Zero
Fixture
.
3.
Press the button and hold the
Zero Fixture
on the probe until the hourglass icon
disappears and a green check mark is
displayed. If the zero fails, a red
X
will
display. Repeat the
Zero
process.
The
Zero
process can take up to 10 seconds
NOTE:
How to Measure
Power-up the
PosiTest
PC
by pressing the button. With the
gage at least 10 cm (4 inches) away from the coated part, press
and hold the button. A target grid will appear on the display:
Target Grid
Distance Target Bar
Distance Paddle
Zero Fixture
Zeroing