4.0 SPECIFICATION
OMD-2005
Range:
0 – 30 ppm, Trend up to 50 ppm
Accuracy
According IMO MEPC. 107(49)
Linearity
Up to 30 ppm better than ± 2 %
Display
Green Graphic Display
Power Supply:
24 V – 240 V AC or DC
Automatic Voltage Selection
Consumption:
< 15 VA
Alarm Points 1 + 2:
Adjustable between 1 - 15 ppm
(Works adjustment 15 ppm)
Alarm 1 Operating Delay:
(for annunciation purpose)
Adjustable between 1 – 540 sec.
(Works adjustment 2 sec)
Alarm 2 Operating Delay:
(for control purposes)
Adjustable between 1 – 10 sec.
(Works adjustment 10 sec)
System Fault Alarm:
Red LED
Alarm Contact Rating:
Potential free 1 pole change over
contacts, 3 A / 240 V
Alarm Indication:
Red LED's
Output Signal:
0 – 20 mA or 4 – 20 mA for 0-30 ppm
reversible, ext. Load < 150
Ω
Sample Water Pressure:
0,1 – 10 bar
Sample Flow:
Approx. 0,1 - 4 l/min depend. to pressure
Ambient Temperature:
+ 1 to + 55° C
Sample Water Temperature:
+ 1 to + 65° C
Roll:
Up to 45°
Size (over all):
360 mm W x 240 mm H x 100 mm D
Degree of Protection:
IP 65
Weight: 7,3
kg
Pipe Connections:
R ¼" Female