DECKMA HAMBURG GmbH
Issue: 27.05.08
Instruction Manual OMD-2005
Page 12 of 25
Signal Output
0(4)-20 mA
To Alarmsystem
3/2 Way Valve
Automatic Stopping Device
Power Supply
Air Supply
Solenoid Valve
1
2
3
19 20 21 22 23 24 25 26 27 28 29 30 31 32 33
Measuring
Cell
Logger /
Display
F2
F1
Res.
PILOT V
24-240V IN
FLOW
STATUS
RES
OUTPUT
CLEAN
PE
PE
PE
+
-
PE
PE
L
N
PE
4
5
6
PILOT OUT
L
N
PE
7
8
9
A
B
PE
RES
13
14
15
NO
COM
NC
ALARM 2
16
17
18
NO
COM
NC
10
11
12
NO
COM
NC
ALARM 1
SYSTEM FAULT
1A
2A
Contacts shown
in Alarm condition
(de-energised)
Status
Separator
Limit
Switch
To Alarmsystem
(optional)
1-2 Power Supply
4-5 Pilot Voltage Output (Same as Power Supply)
7-8 Spare Voltage Output (Same as Power Supply)
10-12 Potential free Output Alarm 1 (Change over contact)
13-15 Potential free Output Alarm 2 (Change over contact)
16-18 Potential free Output System Fault (Change over contact)
19-20 Input Flow Direction Switch (Deckma Delivery)
22-23 Input Status Switch from Separator (Close when running)
25-26 Input Spare Switch
28-29 Signal Output 0(4) to 20 mA
31-32 Optional Output for Autoclean Valve
EXAMPLE
Connections may vary
with different separator
control boxes
Fig. 5
Close front door complete after electrical installation. Water inside the instrument
may result in corrosion and malfunction.