![Decagon Devices ProCheck Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/decagon-devices/procheck/procheck_operators-manual_2461479019.webp)
ProCheck
4. The Menus
15
Setting SDI-12 addresses
The ProCheck may be used to conveniently assign SDI-12
addresses to the Decagon digital sensor. To assign the ad-
dresses, follow the instructions below:
Use the pigtail adaptor to connect a SDI-12 compat-
1.
ible sensor the the ProCheck’s stereo jack.
In the CONFIG menu, select SDI-12 Address. Once
2.
selected the following screen will appear. The screen
will refresh with the sensor’s current configuration.
Press the up and down arrows to select the new ad-
3.
dress from “0” (default) to 1-9, a-z, or A-Z.
Press ENTER or SAVE to accept new address. Screen
4.
will refresh showing new address setting.
Press ESC or MENU to exit the SDI-12 Address Sub
5.
Menu.
Note: If you plan on setting addresses on multiple sensors,
repeat the steps 1 through 4 before proceeding to step 5.