4. The Menus
11
Below is a list of sensors that support soil calibrations.
Sensor Name
Supported Medium Calibrations
EC-20 Soil Moisture
Mineral, Custom
EC-10 Soil Moisture
Mineral, Custom
10 HS Soil Moisture
Mineral, Custom
EC-5 Soil Moisture
Mineral, Potting, Rockwool, Perlite, Custom
ECH
2
O-TE Moisture,
Temp, EC
Mineral, Potting, Rockwool, Perlite, Custom
EC-TM Moisture, Temp
Mineral, Potting, Rockwool, Perlite, Custom
5TE Moisture, Temp, EC
Mineral, Potting, Rockwool, Perlite, Custom
5TM Moisture, Temp
Mineral, Potting, Rockwool, Perlite, Custom
NOTE: With the 5TE 5TM sensors (only), custum calibra-
tion equations must be constructed with
e
a
(dielectric per-
mittivity) as the ‘x’ variable. With the ProCheck and other
Decagon data acquisition equipment,
e
a
= Raw/50. With
other data acquisition equipment, if read in SDI-12 mode,
e
a
= Raw/100.
Creating a Custom Calibration
You can create your own custom calibration for certain sen-
sor types. A custom calibration is applied directly to the