Starting a test // Calibrating
AUTOMATIC MODE
3
Evaluator should press START/STOP
button to Start the test.
Start / Stop
The STATUS (S) will change from STOP
(STP) to CALIBRATING (CAL).
Ts:A
Tr:4
d:0.00
S:CAl
t:0
The STATUS (S) will change from STOP
(STP) to ADJUSTING (ADJ).
Ts:A
Tr:4
d:-90.2
S:ADJ
t:0
a.
If the rod is not at zero position (0) or if not aligned with the fixed rod
indicator, the system will need a self-calibration to zero position (0).
b.
In the case when the distance is zero (0) or the indicators are aligned,
system will not need to calibrate. System will start ADJUSTING (ADJ)
mobile rod distance immeditely.
Warning: Once the equipment has finished the automatic adjustment of the movable rod the
equipment state (S) will change immediately from Adjust (ADJ) to Start (STRT), be sure the
evaluated subject holds the joystick control, as evaluation will begin right at this point.
Starting a test
Time (T) will start running.
Ts:a
Tr:4
d:-87.6
S:STRT
t:2083
t:207O
t:2011
A white light inside the equipment body will
turn on.
Immediately after STATUS (S) will change
from ADJUST (ADJ) to START (STRT), these
two things will happen:
Ts:A
Tr:4
d:-87.6
S:ADJ
t:0
a.
When equipment is set at Automatic (A)
mode, each Trial Location (TR) has a
predeterminated distance for the moving
rod, these are the values for each TR:
TR # 0: -90.2
TR # 1: 1.25
TR # 2: 62.57
TR # 3: 87.6
TR # 4: -87.6
TR # 5: 37.53
TR # 6: -37.53
TR # 7: -1.25
TR # 8: -62.57
TR # 9: 90.2
PROCEDURE
// OPERATING