daviteq WSSFC-V1A-025 Скачать руководство пользователя страница 2

WSSFC-V1A is a cost-effective, single-axis vibration sensor designed for condition monitoring and preventive

maintenance applications. The piezo-electric accelerometer is available in ranges ±25g or 50g and features a flat

frequency response up to >10kHz. Its accelerometer feature a stable piezo-ceramic crystal in shear mode with low-

power electronics, sealed in a fully hermetic package. The Piezo Electric technology incorporated in the WSSCF-V1A

accelerometer has a proven track record for offering the reliable and long-term stable output required for condition

monitoring applications. The accelerometer is designed and qualified for machine health monitoring and has superior

Resolution, Dynamic Range, and Bandwidth to MEMS devices. Besides that, it can also measure the temperature at the

mounting point. 

It is battery-operated and able to connect to any Sigfox network in the World. It supports all frequency zones such as

RC1, RC2, RC3c, RC4, RC5, RC6, and RC7.

For the principle operation of the V1A single-axis vibration sensor, please refer t

this link.

Please refer to 

this link

 for typical applications.

The device will send uplink messages in the following cases:

Case 1: After power-up in the 60s, the device will send the first message called START_UP. The payload will tell

the user the HW version, FW version, and current configuration of the device;

Case 2: Then, in every interval time (pre-configured), for example, 10 minutes, it will send the message called

CYCLIC_DATA. The payload will tell the user the following data like measured values, battery level, alarm status...

Case 3: If the Alarm function was enabled (in the configuration of the sensor), if the measured value passed the

threshold, it will send the uplink message immediately. This message is called ALARM. The payload also tells the

user the data like measured values, battery level, alarm status...

Case 4: The HEART_BEAT uplink message will be sent once a day (the default setting can be changed in

configuration) to allow the Sigfox back-end system can send the downlink message for changing the

configuration of the sensor. Please refer to the downlink section for more details. The uplink payload will tell the

user the HW version, FW version, and current configuration of the device;

Case 5: During commissioning, testing, or calibration sensor, the user can force the device to send the uplink

message to get the data immediately. This message is called FORCE_DATA. The payload will provide data like

raw measured value, scaled measured value, battery level, alarm status... It can be forced by applying the

magnet key on the reed switch in 1s;

Case 6: If users want to change the configuration immediately, they don't need to wait up to 1 day for the

HEART_BEAT message, instead they can force the device to send a special uplink message so that the device can

get the new downlink message. This uplink message is named PARAMETERS_UPDATE. It can be forced by

applying the magnet key in more than 5s.

The sensor was pre-configured at the factory with default values for configuration parameters that meet most use

cases. However, depending on the specific use case, the customer can adjust those parameters. Please refer to

section 3.2

 for more details.

The sensor is powered by 2 x AA 1.5V batteries for many years of operation. We do recommend using Energizer L91

battery which is very popular and high performance. This battery has a capacity of up to 3500mAh with a working
temperature range from -40 to +60 oC. The instruction for installing the batteries is in 

this link

.

Figure 1. Battery Energizer L91

Finish this part so you can understand and put the sensor in operation with the default configuration from the

factory.

1.1 What is the Sigfox-Ready V1A Single-Axis Vibration Sensor
and its principle of operation?

1.1.1 What are the typical applications of this sensor?

1.1.2 When does the device send uplink messages?

To change the cycle of data sending, you can change the value of the parameter: CYCLIC_DATA_PERIOD (default

is 600 seconds).

The alarm thresholds can be changed via downlink or offline tools.

1.1.3 The important configuration parameters

1.1.4 What kind of battery is used for this sensor?

Содержание WSSFC-V1A-025

Страница 1: ...ct Type 2 parts Mounting Type Direct process mounting for sensor wall mount for transmitter Powered by 2 x AA batteries 1 5V Information Changes in this version v s previous version Item Changes Changed by Changed Date Approved by Approved Date 1 Initial version D Q Tuan 29 08 2022 N V Loc 05 09 2022 To use this product please refer step by step to the below instructions Operating Principle Uplink...

Страница 2: ...llow the Sigfox back end system can send the downlink message for changing the configuration of the sensor Please refer to the downlink section for more details The uplink payload will tell the user the HW version FW version and current configuration of the device Case 5 During commissioning testing or calibration sensor the user can force the device to send the uplink message to get the data imme...

Страница 3: ...ey 01 x Wall mounting bracket and screws 01 x Vibration sensor module V1A 1 3 Quick Test Step 1 Prepare the values of communication settings Note All Sigfox sensors are pre configured with the correct RC before delivery The settings of Device ID Device PAC and RC could also be read from the device memory map Please reference section 3 2 Sensor configuration for details Step 2 Add the device to Sig...

Страница 4: ...s transmitter must be mounted at the minimum level of 2m from the ground for a better RF signal To get the strongest RF signal please follow this link How to mount it with a mounting bracket please checkthis guide Insert the batteries into the wireless transmitter and check the system to see whether the wireless transmitter already sent the first message to the system Please follow this link to le...

Страница 5: ...oot the sensor part Maintenance works Yes No Descriptions Consumable parts replacement Yes The battery is the only part need to check the lifetime to replace Check the battery status on the back end system Cleaning device No Re calibration Re validation No No calibration is required for the wireless transmitter Please refer to this link 1 5 3 Device calibration configuration 2 Maintenance 2 1 Trou...

Страница 6: ...r tothis link for a complete understanding of this measuring technique Below are some important configuration parameters which affect the operation of the device like battery life measurement accuracy and alert threshold For Battery life estimation please refer to this link measure_period Default 3600s This is the time period for the wireless transmitter to wake up and take the measurement from th...

Страница 7: ...rameters of the device Read only Parameter Table Modbus Register Decimal Modbus Register Hex Function Code Read No of Registers Description Range Format Property Comment 2 2 3 4 FW_VERSION string Read 6 6 3 2 HW_VERSION string Read 8 8 3 2 DEVICE_ID hex Read Product ID 10 A 3 4 DEVICE_PAC hex Read Product PAC 14 E 3 1 SENSOR_TYPE 1 255 uint16 Read Sensor or Input Type Read Write Parameter Table Mo...

Страница 8: ...SET_X10 0 int16 Read Write Offset adjustment for measured temperature value Please refer to this link Please refer to the detailed specifications inthis link For warranty terms and support procedures please refer tothis link Use cases Case studies White papers END Revision 8 Created Wed Sep 7 2022 9 48 AM by Lộc Vĩnh Nguyễn Updated Thu Sep 8 2022 10 06 AM by Phan Van Luc 3 3 Calibration for Sigfox...