7. Remove the white junction board cover and unplug the fan power cable from the 

junction board.

8. Lift out the fan and the fan deflector.

9. Pull the temperature/humidity sensor up and 

out. 

10.Use a soft brush to clean the white plastic and 

gold mesh of the sensor.

11.Remove all debris from inside the shield and 

fan and wipe the interior surfaces with a damp 
cloth.

12.Remove the screen from the bottom of the 

radiation shield. Wipe it clean, as well as up into 
the interior of the radiation shield. Replace the 
screen.

13.Replace the temperature/humidity sensor. It fits 

one way, into the slots on the side. Route the 
cable up through the channel and replace the fan 
deflector with the cable channels correctly 
aligned with the sensor cable. If a new fan and 
batteries are needed, See “Replacing the Fan 
Motor and Batteries” on page 8.

14.Replace the fan and plug the fan power cable 

back into the junction board. The fan should 
start to rotate. 

Fan Unit

Fan Deflector

Temp/Hum

Sensor Cable 

 Channel

Temp/Hum

Sensor Cable 

 

Fan Unit

Fan Power 

Cable

Junction Board
Cover (removed)

Junction Board

Temperature/

Humidity

Sensor

Содержание 6832

Страница 1: ...Hour Solar Powered Fan Aspirated Radiation Shield Product number 6832 Temperature Humidity Sensor USER MANUAL Davis Instruments 3465 Diablo Avenue Hayward CA 94545 2778 U S A 510 732 9229 www davisne...

Страница 2: ......

Страница 3: ...the impact of solar radiation on sensor readings Tools for Setup In addition to the hardware provided you will need some or all of the following materials Adjustable wrench or 7 16 wrench Drill and 3...

Страница 4: ...e Pull the tabs You should hear the fan come on Before You Install the Temperature Humidity Sensor Choose a Location It is important to remember to choose a location that best represents the environme...

Страница 5: ...e to the pole use the 8 cable ties to secure the uncoiled sensor cable to pole to prevent fraying in wind Mounting on a Post 1 Using four 1 4 x 1 1 2 lag screws attach the mounting bracket to the surf...

Страница 6: ...ris or insect nests several times a year and thoroughly clean as often as necessary A buildup of material inside the shield reduces its effectiveness and may cause inaccurate temperature and humidity...

Страница 7: ...convenient work place such as a table a piece of cardboard or a pickup truck tailgate Note Take note of the cable placement and routing as you disassemble the shield so you can replace them correctly...

Страница 8: ...move the two cap plates Plates Screen Stand offs Fan Plate Temp Humidity Cable Solar Panel Cable Plugged into Junction Board Closed Cap Plate Open Cap Plate hole in center Solar Panel Bracket Junction...

Страница 9: ...p into the interior of the radiation shield Replace the screen 13 Replace the temperature humidity sensor It fits one way into the slots on the side Route the cable up through the channel and replace...

Страница 10: ...rd Motor Kit for Fan Aspirated Radiation Shield with Batteries Troubleshooting I am not receiving data from my sensors Make sure the sensor cable is firmly connected Try reseating the connections My d...

Страница 11: ...Induced Error 4 F 2 C at solar noon insolation 1040 W m2 avg wind speed 2 mph 1 m s Update Interval 10 seconds Alarms High Low Threshold from Current Reading Relative Humidity Resolution and Units 1 R...

Страница 12: ...2 9188 E mail info davisnet com www davisnet com Contacting Davis Instruments If you have questions about your temperature humidity sensor or encounter problems installing or operating the sensor plea...

Отзывы: