8
15. Replace the junction board cover.
16.
R
eplace the two cap plates. (Note that the closed plate goes on top.)
17. Replace the solar bracket, being
careful to use the correct holes in the
bracket as shown.
18. Align the top of the screws with the
holes in the mounting base.
19. Replace the nuts, lock washers and
flat washers.
20. Reconnect the temperature/humidity
sensor cable.
Replacing the Fan Motor and Batteries
To replace the fan motor and batteries in the 24-Hour Fan-Aspirated Radiation Shields, use
product no. 7758B: Standard Motor Kit for Fan-Aspirated Radiation Shield with Batteries.
Troubleshooting
“I am not receiving data from my sensors.”
Make sure the sensor cable is firmly connected. Try reseating the connections.
“My data does not seem to be accurate.”
Make sure the screened opening at the bottom of the radiation shield is not blocked and air
can circulate up the tube. Make sure the fan is running, by listening for it or by placing a
small piece of tissue paper against the screen under the shield.
“The fan is not running.”
Make sure the solar panel cable is firmly connected. If you have changed the batteries, make
sure the fan batteries they were inserted in the right orientation (+ to + and - to -).
USE THESE HOLES
Lock Washer
Hex Nut
Flat Washer