Testing /Placing Into Service
17
Model
Serial No.
=> Note: All the following issues must be considered to complete the reception.
Issue
Status
ISafety notese
Protected individual electric socket-outlet (16 A)
verified
Equipotential bonding
verified
Installation notes / clearance space
verified
Cable routing
verified
Supervision duty
verified
II Functions
ON / OFF switch
verified
EMERGENCY STOP device
verified
Controls mechanical
verified
Controls dashboard / console
verified
III User manual
The user manual was read and understood
.
verified
The user manual must be stored in the direct vicinity of the device
.
=> The reception has been completed in accordance with the Medical
Devices Law (MPG) and the additional directions of the manufacturer.
Received by
Name
Date
Signature
Client
Name
Date
Signature
Reception report
verified