4 Alarms and Troubleshooting
1009-0633-000
4-9
Memory (EEPROM)
Fail
The system cannot access
some stored values.
Default settings are used.
Ventilation is still possible but
service is necessary.
Contact a qualified
service representative.
Memory (flash)
Failure
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Memory (RAM)
Failure
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Memory (Redundant
Storage) Fail
Minimum
monitoring
(Medium)
Ventilator malfunction.
Ventilate manually. Monitoring is
still available.
Contact a qualified
service representative.
Memory (video)
Failure
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Monitoring Only
Medium
A severe malfunction
prevents mechanical
ventilation. Other alarms
may also occur.
Ventilate manually. Cycle system
power (On- Standby-On). If the
alarm clears, restart mechanical
ventilation.
Contact a qualified
service representative.
No Circuit Module
Low
Ventilator malfunction.
Ventilate manually.
Optical sensors look for
tabs on the back of the
module. Is the module
assembled? are the
sensors dirty?
No CO
2
Absorption
Medium
(low after
acknowledged)
The absorber canister is
open (out of the circuit) but
the bypass mechanism
prevents a leak (optional
feature).
User setting. Reinstall the
absorber canister to remove CO
2
from exhaled gas
- - -
No Exp Flow Sensor
No Insp Flow Sensor
Medium
(low after
acknowledged)
Electrical signals show the
flow sensor is not
connected.
Connect the flow sensors. Make
sure the flow sensor module is
on all the way.
- - -
No message, only
specific shutdown
message
High
A severe malfunction
prevents mechanical
ventilation and monitoring.
Other alarms may also
occur.
Ventilate manually. Use a stand-
alone monitor. Cycle system
power (On- Standby-On). If the
alarm clears, restart mechanical
ventilation.
Contact a qualified
service representative
No O
2
Pressure
High
(cannot be
silenced)
The O
2
supply has failed.
Air flow will continue. Ventilate
manually if necessary. Connect a
pipeline supply or install an O
2
cylinder.
- - -
Message
Priority
Cause
Action/Concerns
Repair