Breathing system maintenance
When cleaning the breathing system, replace any parts that are visibly
cracked, chipped, distorted or worn.
Refer to the appropriate section for reassembly and tests.
O
2
sensor replacement
w
WARNING
Handle and dispose of sensors according to your biohazard
policies. Do not incinerate.
1. Pull the latch (1) to unlock the flow sensor module from the breathing
system.
2. Pull the flow sensor module from the breathing system.
3. Remove the O
2
sensor cable connector (2) from the O
2
sensor (1) and
unscrew the sensor counterclockwise. Don’t lose the sensor o-ring.