
Aespire 7900
2-26
1009-0633-000
Mechanical cleaning in
washer or washer-
disinfector
1. Place the canister without filters and the lid in the washer or washer-
disinfector and clean them with a decontaminating program.
2. Dry the canister and the lid in a warming closet, maximum 80°C (176°F)
or at room temperature.
3. If the washer or washer-disinfector is not used for disinfection of
equipment, Datex-Ohmeda recommends that a further high level
disinfection is conducted.
4. When dry, place a new filter in the bottom of the canister, pour soda lime
into the canister and place a new filter over the soda lime before closing
and locking the cover. Wipe off any soda lime dust.
5. Align the cover slots with the canister locking tabs, and press the cover
down into place. Turn the cover locking ring clockwise to lock the cover in
place. Ensure cover is properly sealed to prevent leaks and spillage.
Alignment of arrows helps to indicate correct assembly.
Manual cleaning
1. Flush the canister and the lid with fresh running water.
2. Clean the canister and lid under total immersion in a sink with water and
cleaning agent for at least 3 minutes. The water temperature should be
approximately 40°C (104°F).
3. Flush the canister and lid with fresh running water.
4. When dry, place a new filter in the bottom of the canister, pour soda lime
into the canister and place a new filter over the soda lime before closing
and locking the cover. Wipe off any soda lime dust.
5. Align the cover slots with the canister locking tabs, and press the cover
down into place. Turn the cover locking ring clockwise to lock the cover in
place. Ensure cover is properly sealed to prevent leaks and spillage.
Alignment of arrows helps to indicate correct assembly.
6. Datex-Ohmeda recommends that manual cleaning is always followed by
a high level disinfection.
High level disinfection
1. Always clean the canister before high level disinfection.
2. The canister can be steam autoclaved. Maximum recommended
temperature is 134°C (273°F).
3. When dry, place a new filter in the bottom of the canister, pour soda lime
into the canister and place a new filter over the soda lime before closing
and locking the cover. Wipe off any soda lime dust.
4. Align the cover slots with the canister locking tabs, and press the cover
down into place. Turn the cover locking ring clockwise to lock the cover in
place. Ensure cover is properly sealed to prevent leaks and spillage.
Alignment of arrows helps to indicate correct assembly.
Содержание Aespire 7900
Страница 10: ...Aespire 7900 viii 1009 0633 000 ...
Страница 78: ...Aespire 7900 4 18 ...
Страница 116: ...Aespire 7900 I 2 1009 0633 000 ...