Comment: the precondition for the following examples is
that the output is in
mode, that is, it switches ON when
the measured value
≥
L1.LL (low limit). The input signal
has an increasing characteristic, but rippled.
value
= 1000. The value of the relevant hysteresis (
) = 0.
Example:
Hysteresis > 0
A hysteresis with a value of zero means that the limit
detection has no hysteresis. The output switches ON when
the value on the display reaches or is in excess of
,
and switches OFF when it goes lower than
again.
The following figure shows the essentials of
hysteresis, when it’s value is higher than zero
Figure:
Hysteresis > 0