DT4220 E
8
20180619-V0
Operating principle
The input current or voltage to be measured is led to the 23 bit A/D
converter through a protection and signal condition circuit.
The digital output signal of the A/D converter is processed by a
microcontroller.
The microcontroller drives the 5 digit LED display, processes the front
panel membrane keypad, drives the limit relays, the analogue output
and the communication output.
Easy on-site configuration through the front panel membrane keypad
is a major advantage of the microcontroller-based technology. The
configuration parameters: input signal selection, signal filtering,
display scaling, decimal point position, display refresh rate, limit
modes, limit values, etc. are stored in FRAM. The settings are
protected by a two-level password against unauthorized manipulation.
Power supply
The instrument has two power supply version:
DT4220 E xx xx xx:
19-50 VDC / 19-35 VAC
3 W / 3.5 VA
DT4220 E xx xx xx
PS
:
90-250 V AC/DC 3.5 VA / 3 W
Содержание DT4220 E Series
Страница 1: ...DT4220 E xx xx xx PS Process Indicator Operating Instructions...
Страница 59: ...DT4220 E 20180619 V0 59 Shematic for units of RS 485 communication...
Страница 60: ......