7.5. Two wire zeroing
(03. menu item)
Function
When the Pt1xx / resistance / potentiometer is connected to
the input with two wire, the wire resistance will cause a
measurement error. With this function can be compensated
this error. (The compensation will work only properly at the
same ambient temperature.)
[Factory default: wire resistance = 0 Ohm.]
Note:
At
Pt1000
type instrument
the procedure is the same
but the measured resistance is displayed in one decimal
format.
Sequence of operations
1. Enter the menu with the supervisor code.
The way the code should be type in can be found in
Chapter
7.1. Typing the code in
. You see on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
.(
R
esistance
2
W
ire = 2 wire
resistance mode).
(
Nul
l = zero).
3. Enter the menu item by pressing the
ENTER
button.
4. When the earlier selected wire mode is not 2 wire you will
see:
(
Fail
ed = feiled). The mnemonic is blinking.
The instrument will enter into 2 wire mode
(
Chapter 7.4,
02.
menu item),
here you can select the two wire mode.
5. You will see on the display the measured resistance
value in two (Pt100) or one (Pt1000) decimal format.
6. Shorten the two wire at the sensor side (Pt1xx,
resistance, potentiometer). You will see on the display the
resistance of the connecting wire.
E. g.: the wire resistance is 5.23 Ohm, you will se on the
display:
.
7. If the measured value is stable (the last digit may
change) press the
ENTER
button.
8. You will see on the display:
. The message is blinking
indicating that it can be modify.
9. Pressing the
MENU
button you can exit from this menu
item so the instrument will taken into account the last stored
wire resistance.
Содержание DT4120 PS Series
Страница 1: ...DT4120 xx xx PS Temperature Resistance Potentiometer Meter Transmitter Operating Instructions...
Страница 84: ...DT4120 xx xx PS 84 20170324 V0 10 2 Application example Analog output option...
Страница 85: ...DT4120 xx xx PS 20170324 V0 85 RS232 communication interface option...
Страница 86: ...DT4120 xx xx PS 86 20170324 V0 RS485 communication interface option...
Страница 87: ...DT4120 xx xx PS 20170324 V0 87 RS485 bus topology...
Страница 90: ...DT4120 xx xx PS 90 20170324 V0...
Страница 91: ...DT4120 xx xx PS 20170324 V0 91...
Страница 92: ......