6.2. Indicators
The following figure shows the front panel of the instrument with the
USB connector and indicators:
(2) "error" indicator (LED located on the front panel, under the USB
connector), which indicates the error status with a red, flashing
light.
The sequence number of the error is given by the number of flashes.
(3) "on" indicator, which indicates the power on state of the instrument
with a continuous, green light.
Error status indication:
The “error” indicator: flashes red.
The "error" indicator can indicate the following error states:
•
1 flash
: Serious error
•
2 flash
: Broken input wiring
•
3 flash
: Uncalibrated potentiometer
•
4 flash
: Input signal out of range
•
5 flash
: Output signal out of range