A-Class
79
Step
Action
Displayed Message
Comment
I
Lower and latch the Head
Lift Lever.
Using the UP or the DOWN
ARROW Key, set the Gain
Number to 00 and then
press ENTER.
Record this reading as a
TOF Value for Gain
Number 00 in the table.
GAIN TRAN
*00 (0 - 31)
<yyy>
- Or, for reflective media:
GAIN REFL
*00 (0 – 31)
<yyy>
Where ‘yyy’ represents
the current sensor
reading.
After ENTER is pressed,
an asterisk (*) appears
to denote selection.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252 248
01 250
02 248
… …
31 09
Step
Action
Displayed Message
Comment
J
Use the keys to increment
the Gain Number by one
and then press the ENTER
Key. Record the TOF Value.
Repeat this step for each
Gain Number.
GAIN TRAN
*01 (0 - 31)
<yyy>
- Or, for reflective media:
GAIN REFL
*01 (0 – 31)
<yyy>
Where ‘yyy’ represents
the current sensor
reading.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252 248
01 250 245
02 248 234
… … …
31 09 14
(continued)
Содержание A-Class A-4212
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 14: ...vi ...
Страница 20: ...6 A Class 1 2 1 Right Hand Models 1 2 3 4 6 7 8 9 5 27 28 15 16 17 18 19 14 20 21 10 11 12 13 3 22 23 24 25 26 ...
Страница 21: ...A Class 7 1 2 2 Left Hand Models 1 2 3 6 7 8 9 5 4 27 28 15 16 17 18 19 14 20 21 10 11 12 13 22 23 24 25 26 ...
Страница 22: ...8 A Class ...
Страница 84: ...70 A Class ...
Страница 128: ...114 A Class ...
Страница 136: ...122 A Class ...
Страница 140: ...126 A Class ...
Страница 144: ...130 A Class ...
Страница 158: ...144 A Class ...
Страница 160: ...146 A Class ...
Страница 166: ...152 A Class ...