146
Maintenance
We recommend following the instructions and precautions herein to achieve high reliability of
the programmer for a long period of time.
The programmer maintenance depends on character and amount of its use. Regardless, the
following recommendations are generally accepted:
Do not use and store the programmer in dusty places.
Humidity accelerates sedimentation of debris and dust in ZIF socket.
After end the job cover the ZIF socket.
Do not expose the programmer to direct sunlight or position near a source of heat.
Intensively daily use (programming centre, production)
Daily maintenance
Check the ZIF sockets of the programming module for their condition and wear. Remove
debris, dust and grime from the ZIF sockets with clean, dry and compressed air. Clean the ZIF
sockets both in closed and opened position.
For DATAMAN 448PRO2AP-AU: check a proximity sensors and a rotor of Actuation unit if are
clean. If not, remove debris, dust and grime from them.
Weekly maintenance
Perform the Selftest for every programmer or programming module.
Quarterly maintenance
Gently clean the surface of the programmer with isopropyl alcohol or technical alcohol on a
soft cloth.
Perform the calibration test if the programmer supports this feature.
Daily use (developing laboratory, office)
Daily maintenance
After end of the job cover the ZIF socket of the programming module. It is also recommended
to protect the ZIF socket of programming modules from dust and grime.
Weekly maintenance
Check the ZIF socket of the programming modules for their condition and wear. Remove
debris, dust and grime from the ZIF sockets with clean, dry and compressed air. Clean the ZIF
sockets both in closed and opened position.
Quarterly maintenance
Perform the Selftest for every programmer or programming module.
Biannual maintenance
Gently clean the surface of the programmer with isopropyl alcohol or technical alcohol on a
soft cloth.
Perform the calibration test if the programmer supports this feature.
Occasional use
Daily maintenance
After end of the job cover the ZIF socket of the programming module. It is also recommended
to protect the ZIF socket of programming modules from dust and grime.
Quarterly maintenance
Check the ZIF socket of the programming modules for their condition and wear. Remove
debris, dust and grime from the ZIF sockets with clean, dry and compressed air. Clean the ZIF
sockets both in closed and opened position.
Содержание 448PRO2AP-AU
Страница 6: ...6 Introduction...
Страница 11: ...DATAMAN 448PRO2AP AU 11 DATAMAN 448PRO2AP AU...
Страница 19: ...DATAMAN 448PRO2AP 19 DATAMAN 448PRO2AP...
Страница 26: ...26 DATAMAN 48PRO2AP...
Страница 33: ...Setup 33 Setup...
Страница 38: ...38 Step 8 Installation process will start Step 9 Click Finish button to finish setup...
Страница 56: ...56 PG4UW...
Страница 123: ...PG4UWMC 123 PG4UWMC...
Страница 136: ...136 Installation procedure customized...
Страница 145: ...Common notes 145 Common notes...
Страница 157: ...Troubleshooting and warranty 157 Troubleshooting and warranty...