Calibration Verification (Kilograms)
Product Reference Guide
89
Increasing Load Test (Phase 2 Dual Interval ONLY)
After completing the Shift Test, you must complete the Increasing Load Test
using 4, 5 and 6 kilograms of weight.
Shift Test — Metric (5 Kilograms)
The Shift Test checks to ensure that items placed anywhere on the weighing
surface of the scanner/scale are weighed properly. Refer to
when
performing this test.
NOTE
The upper limit of the scale is configurable according to POS interface type and may
not necessarily be set at 15 kilograms, which is the standard setting. For this test,
continue to place weights in 1 kilogram increments only to the upper weight limit
set for your scale.
DUAL INTERVAL
1. Place 4 kilograms in the center of the weighing
surface and check that the Remote
Display reads between 3.998 kg and 4.002 kg.
2. Place an additional 1 kilogram in the center of
the weighing surface and check that the display
reads between 4.998 kg and 5.002 kg.
3. Place an additional 1 kilogram in the center of
the weighing surface and check that the display
reads between 5.998 kg and 6.005 kg.
4. Increase the weight on the center of the weigh-
ing surface to 7.50 kg and check that the display
reads between 7.495 and 7.505 kg.
5. Remove the weights and verify that the display
reads 0.000 kg. You have completed the
Increasing-Load Test.
SINGLE INTERVAL
DUAL INTERVAL
1. Place and remove in succession, a 5.0
kilogram load on the center of each of
the four quadrants (2, 3, 4, and 5 in
) and in the center (1) of the scan-
ner/scale’s weighing platform. Verify
that the display shows a reading of
between 4.995 and 5.005 kg for each
quadrant/center test and that the dis-
play returns to 0.000 between each load.
1. Place and remove, in succession, a 5.0
kilogram load on the center of each of the
four quadrants: 2, 3, 4, and 5, and in the
center (1) of the scanner/scale’s weighing
platform (see
). Verify that the
Remote Display shows a reading of
between 4.998 and 5.002 kg for each
quadrant/center test. Check that the
display returns to 0.000 between each load.
2.
A
fter verifying the accuracy of each
quadrant and the center of the weighing
surface, remove all weight from the
scale. This completes the Shift Test —
Metric (5 Kilograms).
2. A
fter verifying the accuracy of each
quadrant and the center of the weighing
surface, remove all weight from the scale.
This completes the Shift Test — Metric (5
Kilograms).
Содержание Magellan 9300i
Страница 1: ...Magellan 9300i 9400i Scanner and Scanner Scale with Imaging Technology Product Reference Guide...
Страница 12: ...x Magellan 9300i 9400i Scanners NOTES...
Страница 28: ...16 Magellan 9300i 9400i Scanners NOTES...
Страница 60: ...48 Magellan 9300i 9400i Scanners NOTES...
Страница 72: ...60 Magellan 9300i 9400i Scanners NOTES...
Страница 92: ...80 Magellan 9300i 9400i Scanners Figure 46 EAS antenna EAS Antenna Calibration Switch Cover...
Страница 104: ...92 Magellan 9300i 9400i Scanners NOTES...
Страница 114: ...102 Magellan 9300i 9400i Scanners NOTES...
Страница 136: ...Enter Exit Programming Mode Imaging and CSS Features 124 Magellan 9300i 9400i Scanners NOTES...
Страница 196: ...Enter Exit Programming Mode General Scanner and Scale Features 184 Magellan 9300i 9400i Scanners NOTES...
Страница 214: ...202 Magellan 9300i 9400i Scanners NOTES...
Страница 276: ...Enter Exit Programming Mode Interface Related Features 264 Magellan 9300i 9400i Scanners NOTES...
Страница 292: ...Enter Exit Programming Mode Data Editing 280 Magellan 9300i 9400i Scanners NOTES...
Страница 468: ...456 Magellan 9300i 9400i Scanners NOTES...
Страница 472: ...460 Magellan 9300i 9400i Scanners NOTES...
Страница 480: ...468 Magellan 9300i 9400i Scanners NOTES...
Страница 482: ...470 Magellan 9300i 9400i Scanners NOTES...
Страница 518: ...506 Magellan 9300i 9400i Scanners NOTES...