Register Description
Product Reference Guide
23
LAST_SAMPLE_VALUE Register (RO)
Register
Address
Op.
Default
LAST_SAMPLE_VALUE
0x66
RO
0x0000
It represents the last temperature value read from the sensor (even if not
stored in the log memory). LAST_SAMPLE_VALUE register is forced to 0 after
a reset operation.
SAMPLE_NUM Register (RO)
Register
Address
Op.
Default
SAMPLE_NUM
0x67
RO
0x0000
It represents the number of samples present inside the log memory.
SAMPLES_NUM register is forced to 0 after a reset operation.
SHIPPING_DATE Register (RO - 2 words)
Register
Address
Op.
Default
SHIPPING_DATE_L
0x6A
RO
0x0000
SHIPPING_DATE_H
0x6B
RO
0x0000
The Shipping Date register (expressed in the Unix time format) is used to
store the time when logging was enabled (by RF or, if enabled, by the but-
ton). Default value of SHIPPING_DATE register is 0. SHIPPING_DATE register
is set to 0 after a reset operation.
STOP_DATE Register (RO - 2 words)
Register
Address
Op.
Default
STOP_DATE_L
0x6C
RO
0x0000
STOP_DATE_H
0x6D
RO
0x0000
The STOP_DATE register (expressed in the Unix time format) is used to store
the time when logging was stopped (by RF or, if enabled, by the button).
Default value of STOP_DATE register is 0. STOP_DATE register is set to 0
after a reset operation.
USER_AREA Registers (RW - 28 words)
Register
Address
Op.
Default
USER_AREA_0
0x6E
RW
...
USER_AREA_1
0x6F
RW
...
......
......
..
...
USER_AREA_15
0x89
RW
...
User Area is a set of 28 words registers for user defined data storage.
Содержание DLR-TL001