Histogram Description
20
DLR-TL001
MKT_THRESHOLD_TEMP Register (RW)
Register
Address
Op.
Default
MKT_THRESHOLD_TEMP
0x45
RW
0x0000
This register provides the threshold temperature value for the alarm of the
Mean Kinetic algorithm. If the Mean Kinetic value is greater than the thresh-
old value the alarm bit is set. The temperature in °C is expressed in fixed
point 8.5 notation using the formula (3).
SHL_REF_TEMP Register (RW)
Register
Address
Op.
Default
SHL_REF_TEMP
0x46
RW
0x0000
Shelf life reference temperature expressed °C. The temperature in °C is
expressed in fixed point 8.5 notation using the formula (3).
SHL_Q10_TEMP Register (RW - 2 words)
Register
Address
Op.
Default
SHL_Q10_TEMP_L
0x47
RW
0x0000
SHL_Q10_TEMP_H
0x48
RW
0x0000
This value, represented in floating point 32bits (IEEE-754) notation, is used
to configure the Arrhenius algorithm behaviour. It is defined as the ratio of
the reaction rate constants at temperatures differing by 10 °C from the ref-
erence temperature. It is a unit less number. The map between the floating
point fields and the two registers is reported below:
Sign
Exponent
Mantissa
31
30…23 (8 bits)
22…16 15…0 (23 bits)
SHL_Q10_TEMP_H sig
SHL_Q10_TEMP_H exp
SHL_Q10_TEMP_L
SHL_THRESHOLD Register (RW)
Register
Address
Op.
Default
SHL_THRESHOLD
0x4A
RW
0x0000
Shelf life threshold represents the threshold value for the alarm of the
Arrhenius algorithm. If the SHL_TIME value becomes smaller of the Thresh-
old value the alarm bit is set. It is expressed in hours.