
LANcat
User Manual
Page 9-10
Resistance Calibration
Resistance calibration should be performed on a periodic basis in
order to maintain maximum accuracy. Resistance calibration is
most sensitive to temperature changes. Allow the main unit to
stabilize to the operating environment before performing the
calibration procedure. Calibration should be performed whenever
a different Performance Module is installed.
To calibrate Resistance:
1.
Connect the LANcat to the Remote Unit.
2.
Set the dial to CAL.
3.
Highlight Resistance in the Calibrate menu by using the arrow
keys.
4.
Press CALIBRATE (F4) or ENTER. The Calibrate
Resistance screen will appear. The bar graph across the
bottom of the display will show the progress of the
calibration.
5.
When finished, the screen returns to the CALIBRATE menu.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LANcat Series
Страница 130: ...LANcat User Manual Page 7 36 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 156: ...LANcat User Manual Page 9 12 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 162: ...LANcat User Manual Page 10 6 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 182: ...LANcat User Manual Page A 18 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 186: ...LANcat User Manual Page B 4 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 192: ...LANcat User Manual Page C 6 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...