Datacolor 20D User's Guide008-400666 Rev A
41
System Restart
In the event that it is necessary, you can restart the Datacolor 20D without powering it OFF and ON again.
Navigate to the System Info icon on the Live bar and select it:
Select Restart System
.
The program will ask you to confirm the operation. Click
Yes
to restart or
No
to cancel the
operation.
Instrument Maintenance
The certificate of performance supplied with this instrument is valid for one (1) year under the following
conditions:
It is used in an environment suitable for the instrument type.
It is handled and used with care.
Datacolor strongly recommends that the instrument be serviced once/year by an authorized service
engineer.
Datacolor offers a variety of maintenance and certification services to match your needs. Please contact
your local Datacolor office for more information.
The sections that follow provide detailed instructions for maintaining the instrument and calibration tiles.
These instructions and tips will help to insure that the instrument continues to perform properly over its life.
Caution
There are no user-serviceable parts for this equipment.
Instrument Cleaning
The measurement port should be examined visually for the presence of dust, sample particles, fibers, and
excessive yellowing due to environmental influences
. This inspection should be performed frequently,
especially if you measure loose materials.
Loose materials in the sphere should be removed by a Datacolor-certified technician.
Tile Handling and Cleaning
Handling Tiles
Handle calibration tiles with extreme care. Do not drop them, or scratch the glazed surface.
Always use the edges to grasp the tile.