Shimadzu LC-10/20 Pump
3 Installation Procedure
3 Installation Procedure
3.1 Shimadzu LC-10/20 Pump communication
The
Shimadzu LC- 10/20 Pump
is controlled by serial (RS232)
communication. The communication is provided via the Optical to USB
convertor connected to the computer via USB A-A cable (p/n SK10) and to
the pump via optical cable (p/n SK15). Both of these cables are part of the
DataApex Opt-USB
converter set (p/n OPT01), but can also be ordered
separately.
Optical to USB converter
For communication an OPT-RS232 converter is needed. It can either be
the
DataApex Opt- USB
converter (p/n OPT01) or
Shimadzu
adaptor
(Shimadzu p/n 22-35480-91).
Pump Settings
Communication parameters should be set on the pump in the
SERVICE
MODE
(see pump documentation) as follows:
Baud rate
2400
, parity
N
, bits
8
, stop bit
1
.
SYS (Set System Parameter) has to be set to
5
for isocratic mode or to
6
for gradient mode from the pump keyboard.
Caution:
These values are not documented in the pump manual.
The other Pump system parameters should be set to:
Local=0 (Remote)
ADRS= any value (not used)
CBM LINK=0 (not used)
Shimadzu LC-20 settings
Shimadzu LC-20
pumps have to be set to the
LC-10
emulation mode in
order to be able to work in
Clarity
. This is done from the pump's keyboard
according to this procedure:
l
Press the
VP
button on the pump's front keyboard until the
CALIBRATION
inscription is shown on the pump's display.
l
Press the
func
button to enter the pump's configuration. The password
has to be filled in.
Note:
The default password is
00000
.
l
Press the
func
button repeatedly until the
OP MODE
inscription is shown
on the pump's display.
l
Enter the desired number (
1
) to switch the pump to the
SCL- 10Avp
emulation mode and press the
enter
button.
- 3 -