Shimadzu GC-2014C
4 Using the control module
21
options relevant to the detector type selected in the
Method
Setup - GC
dialog.
Signals tab
Fig. 18 Shimadzu GC-2014C Setup - Signals
Shimadzu GC-2014C Setup - Signals
tab defines the number of
signals that will be processed by
Clarity
. These signals will
be assigned to particular detectors (hardware) in the
Method
Setup - GC - DET X
dialog.
Count
Defines the number of signals that will be transferred to the
Clarity
. These signals correspond to
Clarity
detectors and
will have their own sub-tabs in the
Method Setup - Acquisition
dialog, if configured on the given
Clarity
Instrument.
Ch1(..4)
Sets the name of the particular signal.
Enable to start device(s) from Clarity
This checkbox enables to start the analysis run from the
Clarity. Independently on this checkbox, the analysis can be
started automatically by the correctly wired autosampler
after the injection or by the
Start
button on the front panel
of the chromatograph.