Principles of Operation
101
Digital Output Lines
TEMPpoint, VOLTpoint, and MEASURpoint instruments feature eight, latched and isolated
digital output lines. The outputs are solid-state relays that operate at ±30 V and 400 mA peak
(AC or DC). Switching time is 2 ms maximum.
shows the digital output circuitry.
Figure 35: Digital Output Circuitry
Digital outputs resemble a switch; the switch is closed if the state of the digital output line is 1,
and the switch is open if the state of the digital output line is 0. On power up or reset, the
digital outputs are disabled.
Channel-to-Channel Functional Isolation
Each TEMPpoint, VOLTpoint, and MEASURpoint instrument includes channel-to-channel
functional isolation of up to 60 V between digital I/O lines.
Resolution
Each TEMPpoint, VOLTpoint, and MEASURpoint instrument provides 8 bits of resolution for
the digital input port to accommodate the 8 digital input lines and 8 bits of resolution for the
digital output port to accommodate the 8 digital output lines. These lines are organized as
isolated, dedicated ports. You cannot configure port resolution through software.
Operation Modes
Using software, you can read from a single digital input line or the entire digital input port, or
write to a single digital output line or the entire digital output port. You can also return the
value of the entire digital input port in the analog input data stream if you want to correlate
analog input data with digital events; refer to
for more information.
1 of 8 Solid State Relays
Содержание DT9871
Страница 4: ......
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Hardware Features 14 Supported Software 19 Accessories 21 Getting Started Procedure 22 ...
Страница 23: ...Part 1 Getting Started ...
Страница 24: ......
Страница 50: ...Chapter 4 50 Figure 16 2 Wire RTD Connection RTD Channel RTD 425 A Current Sense Sense Return RL RL 1 2 4 3 ...
Страница 80: ...Chapter 5 80 ...
Страница 81: ...Part 2 Using Your Instrument ...
Страница 82: ......
Страница 83: ...83 6 Principles of Operation Block Diagrams 84 Analog Input Features 89 Digital I O Features 100 ...
Страница 102: ...Chapter 6 102 ...
Страница 103: ...103 7 Troubleshooting General Checklist 104 Technical Support 106 If Your Instrument Needs Factory Service 107 ...
Страница 108: ...Chapter 7 108 ...
Страница 129: ...129 B Connector Pin Assignments ...