User Guide
|
Danfoss Gas sensor. Type DGS. Modbus or Service Tool display operation
© Danfoss | DCS (ADAP-KOOL®) | 2018.12
RS8KD102 | 14
4.8
Operating Data
Operating Data
Software version
XXXX
Software version of the measuring point
Operating Data
SC X
Operating days
expected XXX
This menu is for retrieving relevant operational data of the sensor head.
No changes or modifications are possible.
Operating Data
SC
Expected lifetime (in days) of a new
measuring point
Days of Operation
X Days
Current days of operation
Min. Temperature
25 °C
Display of the lowest temperature
detected at the device
Max. Temperature
25 °C
Display of the highest temperature
detected at the device
Maintenance Days
remaining XXXX
Days until the next calibration is due
Operating Days last
XXX
Days of operation when the last
calibration happened
If more than one sensor head is connected to
the DGS, the selection is done at X.
Number of Calib.
X
How many times the measuring point
has been calibrated
Sensivity
XXX%
Remaining sensitivity of the sensor. Must
be replaced when at 35%