56
Falcon2 4M, 8M, and 12M Camera User's Manual
03-032-20107-03
Teledyne DALSA
N ame
flatfieldCalibrationPRN U
Disp lay N am e
Gain(PRN U) Calibration
N am e Sp ace
DFN C
Firm w are Release
00
Visibility
Exp ert
Access
Read -Write w hen in Calibration Mod e
Typ e
Com m and
N otes
Perform s p hoto resp onse non-u niform ity (PRN U) calibration.. PRN U calibration elim inates the
d ifference in resp onsivity betw een the m ost and least sensitive p ixel, creating a u niform
resp onse to light. See the Gain (PRN U) Calibration section in Ap p end ix D: Internal Flat Field
Calibration Algorithm s.
N ame
flatfieldCalibrationGainMode
Disp lay N am e
Gain Calibration Mode
N am e Sp ace
Cu stom
Firm w are Release
05
Visibility
Gu ru
Access
Read -Write w hen in Calibration Mod e
Typ e
Enu m eration
Valu es
HighGain
- Calibrate u sing high gain and low er resolu tion
HighResolution
- Calibrate u sing high resolu tion and low er gain
N otes
Selects the flatfield gain m od e that w ill be u sed in calibration.
H igh gain m od e can ap p ly a correction gain betw een 1 and 2 w ith red u ced resolu tion.
H igh resolu tion can ap p ly a correction gain betw een 1 and 1.5 w ith m axim u m resolu tion.
N ame
flatfieldCalibrationSave
Disp lay N am e
Save Calibration
N am e Sp ace
DFN C
Firm w are Release
00
Visibility
Exp ert
Access
Read -Write w hen in Calibration Mod e
Typ e
Com m and
N otes
Saves the cu rrent flat field coefficients in the Active Set.
N ame
flatfieldCoefficientsCopySource
Disp lay N am e
Copy Source
N am e Sp ace
DFN C
Firm w are Release
00
Visibility
Exp ert
Access
Read -Write w hen in Calibration Mod e
Typ e
Enu m eration
Valu es
FactoryFlatfield
- Factory Calibrated flatfield .
UserFlatField1
to
UserFlatField4
- User configu rable flat field sets.
N otes
Selects the flatfield coefficients set to cop y to the cu rrent Active Set
N ame
flatfieldCoefficientsCopyInCurrent
Disp lay N am e
Copy Coefficient to Active
N am e Sp ace
DFN C
Firm w are Release
00
Visibility
Exp ert
Access
Read -Write w hen in Calibration Mod e
Typ e
Com m and
N otes
Cop ies the cu rrently selected flat field coefficients in the Active Set.