![Dali TE-W2 Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/dali/te-w2/te-w2_user-manual_504147010.webp)
10
[1] Power key
Press and hold the Power Key for > 3 seconds to power on/off the camera.
! Note
: After powering off, please wait ten seconds before powering the
camera on again.
[2] Select/Auto key (Marked “A”)
The “A” or attribute button performs 2 functions.
a) The first function is modification of selected parameters. It is obtained
by a quick press and release (less than 2 seconds), with this functions to
modify their parameters. Subsequent presses will “scroll” through the
available parameters. The selected function will be highlighted in yellow. The
functions include:
Measurement points, areas, and lines
Points, areas, lines – switching the added points, areas and lines in the
interface from Up to Down, press Menu to set properties, press C to delete the
Selected measuring objects.
Color Palette
Pressing the left or right arrows will scroll through the available color
palettes, displaying their name above the palette. The selected palette will
become active after about 3 seconds of selecting it and will be set as the
default power on palette.
Level/Span values
Pressing the left or right arrow will reduce or increase the Span and
pressing the up or down arrow will reduce or increase the Level. If the unit is
in Automatic mode, modifying any of these parameters will select Manual
mode.
b
)
The second function performed is to force an internal calibration cycle.
This is performed by pressing and holding the
A
key for 5
seconds or more.
The camera will make automatic correction to get the most accurate thermal
measurement.
c) Image transmission via blue tooth. Open blue-tooth, select object image,