LT3-P/LT7-P Thermal Image Camera User Manual
DALI
28
4.4.3
Analysis setup
Press
Menu
key to activate the main menu. Select
Setup
->
Analysis
setup
and press
Confirm
key to pop up the Analysis dialog box. User can set
temperature alarm, alarm temperature, alarm color, isotherm color, isotherm
temperature and isotherm width. Press
Up
and
Down
keys to choose menu
items,
Left
and
Right
keys to modify the item contents.
An alarm is triggered when a measurement cursor’s measurement meets
or exceeds the
Set Alarm Temp
setting. All pixels above this temperature will
change to the color as set in
Alarm Color
. When the
Alarm Color
is set to
None
, only the audible alarm will sound. When using alarms with
Area
measurements, the selected measurement mode,
Min
,
Max
or
Average
will
have to meet or exceed the set temperature in order to trigger the alarm.
Temp Alarm
:
Off
disables alarm function,
On
enables alarm function
and allows setting of
Set Alarm Temp
and
Alarm Color
.
Alarming Temp:
Sets the alarm detection temperature. Only enabled
when Temp Alarm is set to
On
.
Alarming Color:
Selects the color in which the alarming pixels are
displayed. If set to None, no changes are made to the screen’s color.
Only enabled when
Temp Alarm
is set to
On
.
Isotherm Color
:
Activates the Isotherm display by displaying the