32) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
33) Repeat steps
31 & 32
until the
VELOCITY
value is correctly displayed.
34) Press the
key to display the
CAL DATA
group.
Note: Since the last data point taken was
ULTRASONIC 2
, or the 6” cal bar,
press the
UP
,
Down
,
Left
,
Right
arrow keys multiple times using the
left button wheel to highlight
A2
for the first reading of the 6” cal bar.
35) Press the
ENTER
key to store the measurement in location
A2
.
36) Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
CAL
menu is highlighted and displaying the
submenu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys
multiple times using the left button wheel.
37) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
MEASURE ZERO
is
highlighted.
38) Press the
ENTER
key to display the
Two Point Zero
list.
39) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
ULTRASONIC 1
is
highlighted.