MAX II
Bolt Tension Monitor
59
9) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
10) Repeat steps
8 & 9
until the
PHY LENGTH
value is correctly displayed.
11) Press the
OK
key to return to the
One Point Zero
list, or
ESC
to cancel
entering the
PHY LENGTH
.
12) Press the
ENTER
key to display the
Digits Edit Box
.
13) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
14) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
15) Repeat steps
13 & 14
until the
TEMP COEF
value is correctly displayed.
16) Press the
OK
key to return to the
One Point Zero
list, or
ESC
to cancel
entering the
TEMP COEF
.
Note:
The
TEMP COEF
should be set to 55 (99 metric) for the 3 & 6” bars
supplied by Dakota Ultrasonics. If the glass block is used, it should be set
to 0. The reason for this is because glass has a very low temperature
coefficient.