PVX-B
Precision Visual Thickness Gauge
29
Selecting Measure Mode - Tabbed Menus
Note:
The
E-E
mode should also be selected for setting up the
PVX-B
in
I-E
mode.
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
GATE
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
MEASURE MODE
is highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys to scroll the mode options. Once the
mode is displayed, press the
MEAS
key to return to the measurement mode
screen.
6.2 Probe zero
The probe zero feature should only be used if a contact style transducer is connected
to the
PVX-B
in pulse-echo mode. As noted in chapter 3, the probe zero function is a
very important and necessary function that must done prior to calibration. It should
also be done on a regular basis. If the
PVX-B
is not zeroed correctly, all of the
measurements taken may be in error by some fixed value. In order to perform a
probe zero, the
PVX-B
must be in pulse-echo mode. If echo-echo mode was
previously selected in the last section, skip this section and continue on to the next.
Note:
If you intend to switch between measurement modes simultaneously, setup
the
PVX-B
in pulse-echo mode, perform the probe zero as necessary, and save the
setup to one of the empty setup locations. Once this is completed, the user can
switch between modes toggling between two setups. The following steps outline the
procedure for performing a probe zero: