background image

DAIHAN-Sci.com or 

ALL

for

LAB.com

Shopping Mall

DAIHAN Scientific, Digital Hotplates & Stirrers

                                                                                                              12/44

Please wear the protective equipment because the heating temperature can be 
overheated above the setting temperature.

Do not operate this product near a heater or an air-conditioner.

Only operate unit after checking the safety of the unit.

If unexpected sound, smell or smog is generated, pull the main plug out from the 
wall socket and contact the technical assets of DAIHAN Scientific.

Do not use outdoors avoid overheating or electrical faults. The use in the place 
where the sun continuously shines and can get rained on it is prohibited. It may 
cause overheating or electrical fault of the product.

Please do not use flammable or volatile products in the vicinity of the product to 
minimize the risk of firer. Please do not use the flammable spray such as lacquers 
and paint, or volatile, flammable chemicals and its similar product, in the 
proximity of the product. They may cause the ignition due to the spark of switch.

Please avoid the disassembly or modification of this product by a non-authorized 
person to avoid serious electrical problems and to retain the warranty service of 
DAIHAN Scientific Co., Ltd.

This product must be connected to a properly grounded power supply.

Please regularly check whether dust or dirt are on the power plug and completely 
connect the plug. If the plug is covered with dust, or the connection is not 
complete, it will cause an electrical shock or fire.

Please separate the power plug from the outlet for safety when the product is 
unused for long periods of time to avoid overheating or ignition due to dust.

When disconnecting the power plug, it must be grasped by the hands and then 
pulled. Grasping only the cord and then disconnecting it may cause overheating or 
ignition because of internal lines being disconnected.

Insert the plug securely into an appropriate wall socket.
Loose connections result in excessive heat generation to the device and may 
cause arcing to occur at the connection.

If unusual noise or signs of abnormal behavior are found during use of this 
product, please immediately contact the supplier or DAIHAN Scientific (Co., Ltd.) 
after turning off the power supply and disconnecting the power line.

Содержание MSH-20D

Страница 1: ...ertified PL Insurance Patented No 10 0554098 Jog Dial Control System User Manual Version2 2 1 written for MSH 20D 30D HP 20D 30D MS 20D Models MSH 30D Digital Hotplate Stirrer DAIHAN Sci com or ALL fo...

Страница 2: ...20D to which this declaration relates is in conformity with the following standard or other normative document EN 61326 1 2006 EN 61010 1 2001 2nd Edition EN 61010 2 010 2003 EN 61010 2 051 2003 foll...

Страница 3: ...D HP 30D to which this declaration relates is in conformity with the following standard or other normative document EN 61326 1 2006 EN 61010 1 2001 2nd Edition EN 61010 2 010 2003 EN 61010 2 051 2003...

Страница 4: ...Product Introduction 6 1 3 Product Configuration 7 2 Installation 2 1 Product Installation 8 2 2 Product Specifications 9 3 Product Usage 3 1 Product Usage 11 3 2 Names of Each Part 14 3 3 Usage Inst...

Страница 5: ...arning Failure to follow the instructions may result in injury Caution Failure to follow the instructions may result in product failure Proximity Caution Keep clear of the product during use Notificat...

Страница 6: ...erformance the lowest noise and its long life 3 Since the precise control of turning rate has been implemented the consistent stirring is provided also at any rate 4 Because white ceramic top plate ha...

Страница 7: ...ce Admission DAIHAN Scientific Co Ltd is a member of Product Liability Insurance PL Insurance so that its customers can trust our products believe and use its products by preparing against the acciden...

Страница 8: ...nic solvents dust or corrosive gases to enter the product 10 Please be careful for burns because the residual heat may be remained even if this product is operating or the power was disconnected 11 Pl...

Страница 9: ...cy Max 380 0 3 at Set Temperature Temp Resolution 0 1 Display 0 5 Control Temp Uniformity Specially Designed Heating Module Temp Difference 10 less Heating Power Consumption 600 W 1 2 kW Timer Alarm 9...

Страница 10: ...206 X 307 X h99mm 286 X 387 X h99mm Weight 2 8 kg 3 3 kg Packing Size 400 X 280 X h200mm 430 X 320 X h200mm Gross Weight 5 kg Power Supply Cord Plug 1 Phase AC 120V 60Hz or AC 230V 50 60Hz Model MS 2...

Страница 11: ...the power cord or the power cord is inserted in between objects the power cord can be damaged and may cause fire or electric shock Please do not directly spray water on the product or wash it with wat...

Страница 12: ...se avoid the disassembly or modification of this product by a non authorized person to avoid serious electrical problems and to retain the warranty service of DAIHAN Scientific Co Ltd This product mus...

Страница 13: ...y removing the foreign matter while gently wiping with the neutral detergent and a dry cloth after checking whether the foreign matter present or not before and after using this product Please always...

Страница 14: ...r cord Power switch Equipment power ON OFF External sensor connector Connection piece of External sensor Ceramic coating top plate Heating Stirring operation Controller Controlling and checking the op...

Страница 15: ...ing Mode LCD Display Input and operation information are displayed Heating Lamp Lights up when the heating operation Stirring Lamp Lights up when the stirring operation Timer Lamp Lights up when the t...

Страница 16: ...nob and if the setting value of the stirring speed is above 80 RPM the value is changed by Turning of the Jog Dial Knob and the START operation is performed by the Push button If turning the Jog Dial...

Страница 17: ...ile blinking the HEATING indication of LCD Display The maximum heating temperature is 380 The Start Stop of the operation is performed by the Push button of Jog Dial Knob The HEATING indication of LCD...

Страница 18: ...ure including the compensation value When entering into the compensation value setting mode during the Heating operation the Heating operation that is operating is stopped After setting the compensati...

Страница 19: ...and held for 2 3 seconds the setting value is saved and this saved value is maintained even when the power is turned on again MODE Locking Mode If pressing once the MODE button in the Set Timer Mode i...

Страница 20: ...Contact Temp Probe SS110 PT 100 Teflon Coated for MSH 30D Probe 4 X L250mm Cable L190mm DH WHP503030 External Direct Contact Temp Probe SS200 PT 100 for MSH 20D Probe 4 X L250mm Cable L190mm DH WHP503...

Страница 21: ...p from a professional CLEANING 1 When cleaning the product wipe with a soft towel after removing the power plug and please wipe only the contaminated area by coating the solvent with low freezing poin...

Страница 22: ...cess about all functions before shipment 2 This instrument is warranted for 2 years from purchasing date 3 A breakdown caused by customer s improper use or natural disaster is excepted from this warra...