IX2-700 Dual Intracellular Preamplifier
11
SPECIFICATIONS
Electrode Holders:
Low noise polycarbonate in two sizes: 1.0 to 1.2mm
(#HBI2O); 1.5 to 1.8 mm (#HB18O) (order separately)
Headstage:
Box Probe, 1.5” x 1.2” x .88”, 4” long, 3/8” rod, BNC
Input Connector
Standard Headstage: N=.l for 20 to 500 Megohm
electrode
Current
Multiplier
Feedback
Resistor
Noise
Electrode
Resistance
Input
Resistance
(N) Max
I
R
E
RMS/BW/Re Range Cp
10 10µA
1M
Ω
25µv, 10kHz, 1M
Ω
.2 to 5 M
Ω
20
pf
10
10
M
Ω
1 1µA
10M
Ω
50µv, 10kHz, 10M
Ω
2 to 50 M
Ω
10
pf/35
pf
10
11
M
Ω
.1(Standard) 100nA 100M
Ω
50µv, 1kHz, 100M
Ω
20 to 500 M
Ω
10
pf/35
pf
10
12
M
Ω
.01 10nA
1G
Ω
40µv, 100Hz, 1G
Ω
200 to 5000 M
Ω
20
pf 10
13
M
Ω
.001 1nA
10G
Ω
35µv, 10Hz, 10G
Ω
2 to 50 G
Ω
20
pf
10
14
m
Ω
Extracellular Recording:
Works with 8024 voltage, gain 1000 AC
coupled headstages
Filter Response:
Switch selectable DC to 1kHz, 3kHz, 5kHz,
10kHz, 30kHz and Wide Band
Input Bias Current:
1 x 10
-
12
Amps (Adjustable to zero)
Rise Time to Pulse:
Typically 10 microseconds into 10 megohm
resistor (10 to 90%), 50 microseconds into
100 megohm
Voltage
Gain: X10
and
X100
Output Impedance:
500 ohms all outputs
Offset Voltage:
+200mV
Содержание IX2-700