5-6. Function List detailed information.
Decimal Point Setting
F01
●
0
No Decimal point
1
1
st
place under Zero (0.0)
2
2
nd
place under Zero (0.00)
3
3
rd
place under Zero (0.000)
Back up mode selection
F02
●
0
Normal mode
1
Back up mode
※
Normal Mode
When the power is off and will not be recovered within 1sec, weight memory will be out and display
wrong weight value when power is recovered.
Also, if there is material over 10% of max. capacity weight in the hopper, the “Un-Pass” display will be
appear and you cannot process weighing job. Under this case, please remove material and turn on
DN511A.
If there is material less than 10% of max. capacity weight in the hopper, indicator will initialize
with ”Pass” display, and display “0” whatever the real weight. Under this case, remove the material
and press “Zero” key and input new “Zero” value. Then, new Zero value will be memorized and you
can process weighing job. For “Manual initialization”, press “0” key or enter “Calibration mode”.
※
Weight Back up Mode
Under this mode, DN511A memories “Zero” value of weighing part. So, when the power is recovered
DN511A can display material weight.
Motion Band Range setting
F03
5
0
+
9
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and
reduce the vibration effect on weighing process.
0 : Weak vibration
+
9 : Strong Vibration
※
This function is compensate vibration effect on weighing part with acceptable range setting.
During the fixed time period(F06 setting), there is smaller weight variation than Motion band range,
due to vibration, Indicator will display “Steady”, if there is larger weight variation than Motion band
range, due to vibration, Indicator will its weight and “Steady” condition is broken and find new “Steady”
point.
If there are much vibration effect, please set with large set value.