Step 3. Max. Capacity of Load cell
(
display)
Input Max. Capacity of Scale with No. keys.
- Under this step, input Total sum of each load cell’s Max. Capacity. (Not weighing Scale)
- The Max. Capacity of load cell is stated on
“Test report”
or
“Label”
.
- If you installed 4 load cells, and each load cell’s Max. Capacity is 500kg, then you have to input
2,000kg, as a Max. Capacity.
Input Capacity and press
key, and move to next step.
Step 4. Measure/Adjustment optimal Zero balance of Scale (
display)
Make empty the scale part, and press
key.
Indicator check the current Zero balance and save the value and move next step.
CAL _0_
XXXXXX