Model CA-1001 Manufacturer’s Declaration – Electromagnetic Immunity
The
Model CA-1001
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Model CA-1001
system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment/Guidance
Conducted RF
3 Vrms
3 V
IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF
3 V/m
3 V/m
IEC 61000-4-3 80 MHz to 2.6 GHz
11 Specification Sheets - Continued
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be considered.
If the measured field strength in the location in which the
Model CA-1001
is used exceeds the applicable RF compliance
level above, the
Model CA-1001
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
Model CA-1001
.
. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Portable and mobile RF communications
equipment should be used no closer to any
part of the
Model CA-1001
, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
=1.2√
=1.2√
80 MHz to 800 MHz
=2.3√
800MHz to 2.3 GHz
Where
is the maximum output power rating
for the transmitter in watts (W) according to
the transmitter manufacturer and
is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
DOC# R01-0006-00004-AA