CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing to be cleaned
and disinfected.
Cleaning and Disinfection
These processes should be used in the handpieces.
Uncouple the piece from the INTRA fit terminal.
Remove the diamond tip or bur.
Clean the external surface with neutral soap and using a clean and soft cloth.
Do not use microabrasive material or steel wool when cleaning, or employ organic solvents
ordetergents which contain solvents such as ether, stain remover, etc.
Spray lubrication
To carry out the lubrication of your equipment, we recommend the use of the Lubs product.
Lubrication must be carried out by the posterior hole of the counter angle. Insert the adapter into
the hole as illustrated in fig. (a).
Firmly fit the oil applicator nozzle into the adapter hole (b) and apply the spray for 1 or 2 seconds,
by observing the lubrication into the instrument head.
Cover the piece with absorbent paper to remove the excess of oil.
Warning
Never submerge the instrument in a disinfection bath.
Never use product peracetic acid for cleaning, since it is corrosive.
Lubricating
Adapter
Warning
Before lubricating the handpieces, remove the bit.
fig. a
fig. b
The lubricant doesn’t follow the product.
Attention:
The non-use of the lubricant would result
in the loss of the product warranty.
NOTE: Handpieces not subjected to sterilization process should be lubricated at least twice
a day.