Principles and fundamentals applied to the product functioning

Sterilization by autoclave consists of exposing the contaminated material to steam under high 

pressures and temperatures, for the time needed. Humidity and heat irreversibly denature structural 

enzymes and proteins, destroying microorganisms.

The process includes compression and decompression cycles as to provide easier contact between 

the steam and the contaminated material. The pressure value ranges from 1.3 to 2.1 kgf/cm2 and 

temperatures from 123°C to 133°C.

*

 Optional

Description of Equipment

It’s an autoclave for sterilization by vapor, modern design, easy to operate, ideal for dentistry offices

(general practice, endodontia, pediatric dentistry, orthodontics, prosthesis, operative dentistry and

esthetics).

Its structure is made in carbon steel, with surface treatment through nanotechnology.

Smooth high gloss epoxy based paint polymerized at 250ºC with corrosion resistant treatment and 

cleaning materials.

Its chamber is made of stainless steel

*

 or aluminum

*

, offering long durability and resistance to the

constant variance of temperature, with capacity for 12 Liters.

It’s internally coated with heath isolating material, which optimizes its operation and reduces the

energy consumption.

It features an anti-vacuum valve, which prevents the door and safety seal from blocking, and it’s 

automatically activated if an excessive pressure was detected. The door’s sealing system is made of

high-performance silicone, resistant to temperature and pressure changes, and its flexibility allows 

the perfect sealing between surfaces in contact, preventing leakage.

The control panel contains LED and function buttons, graduated manometer for temperature and 

pressure, ergonomic handle with automatic depressurization safety system, which prevents accidents

in case of unintentional opening.

Accompanied by a hose for outlet of steam, graduated cup, spare fuses and 2 trays made of stainless 

steel or aluminum totally pierced to allow a good circulation of heat, ensuring excellent quality 

in the sterilization.

Basic Principles of Autoclave’s Operation:

• Water supply for autoclave.

• Heating: Water steam followed by gradual increase of pressure.

• Sterilization: High pressures and temperatures through approximately 16 minutes.

• Depressurization: Steam release and relief of autoclave pressure.

• Drying: Material drying and cooling until achieving a temperature which allows handling.

IDENTIFICATION OF EQUIPMENT

Содержание AUTOCLAVE BIOCLAVE 12 LITERS

Страница 1: ......

Страница 2: ...Est 582 002 897 114 Rod Abr o Assed Km 53 450m Cx Postal 782 CEP 14097 500 Ribeir o Preto S P Brasil Phone 55 16 3512 1212 Technical Duties Ricardo J Ravaneli CREA SP 5060714523 Registration ANVISA n...

Страница 3: ...cleaning and disinfection of equipment 22 Precautions in case of alteration in the functioning of equipment 22 Precautions to be adopted against foreseeable or uncommon risks related to the deactivati...

Страница 4: ...ion of the country and following the instructions of the manual The operation of the equipment required for the professional the utilization of correct instruments and it should to be in perfect condi...

Страница 5: ...12 Liters It s internally coated with heath isolating material which optimizes its operation and reduces the energy consumption It features an anti vacuum valve which prevents the door and safety sea...

Страница 6: ...y between equipment coupling and accessories 01 Autoclave s body 02 Front door 03 Manometer 04 Control panel 05 Door latch 06 Plug input power cable 07 Voltage selector switch 08 Fuse holder 09 Master...

Страница 7: ...hose This accessory must be connected and fixed to the external steam outlet using a clamp It takes steam out 03 Graduated glass It s used to dose the quantity of distilled water needed for each steri...

Страница 8: ...r in case it exceeds The anti vacuum valve has the same function and it also reduces the pressure of the chamber if it were negative vacuum NOTE If during the sterilization procedure the valves were o...

Страница 9: ...from entering the steam outlet External steam outlet The 5 16 hose fitting found in the autoclave rear part takes the steam outside Model Autoclave 12L Power Supply Voltage 127V 100 140V 220V 198 264V...

Страница 10: ...23 Kg 26 Kg Fuse of Protection F1 0 5A F2 and F3 15A F4 08A Full cycle time 60 minutes These values are approximate they may change depending on the material s type and volume energy oscillation etc M...

Страница 11: ...11 Dimensions mm For installation of the Autoclave in niches cabinets consider the addition of 50mm in the total length of the product TECHNICAL SPECIFICATIONS...

Страница 12: ...irect sun light exposure Packing to be transported and or stored with care should not suffer drop and neither receive impact Temperature limit for the packing to be stored or transported Applied part...

Страница 13: ...llation and Maintenance Manual of the equipment that is found with the Alliage authorized technical representative Important safety considerations Never touch the steam external outlet and after the s...

Страница 14: ...tected from current overload by fuses Special precautions and installation conditions Install the autoclave on a firm and leveled surface approximately at 80 cm from the floor Leave enough space for h...

Страница 15: ...t is advisable to always have spare fuses at hand All fuses must be installed independently from the grid voltage Preparing the autoclave Connect and fix the rubber hose 10 using the clamp 11 to the e...

Страница 16: ...fore first use please remove the protector film which wraps this tray 27 The female outlet should be a grounded bipolar type three prong 2 flat and 1 round Verify that the ground wire is properly conn...

Страница 17: ...ave s door pushing it against the tank pushing the door latch to the right and then lowering it Check that it is correctly closed 23 In order to avoid accidents and burns it s very important that the...

Страница 18: ...y depressurized before opening the door 10 In case of power cut the operation of the autoclave will be interrupted the yellow 4G LED will come on and it will remain blinking activating the discharge v...

Страница 19: ...it Open the autoclave s front door and wait until it s cold Proceed as described in the drawing pressing them until they go back to their original position If the valves were opened repeatedly call a...

Страница 20: ...tmospheric pressure range 700 hPa to 1060 hPa 525 mmHg to 795 mmHg Operation altitude 2000m Attention The Equipment maintains its condition of safety and efficacy provided that it is maintained stored...

Страница 21: ...e equipment which could cause short circuits Never touch the steam external outlet and after the sterilization is finished never touch the internal autoclave s surfaces The inappropriate use may cause...

Страница 22: ...after the use of equipment Turn off the main switch of the dental set when it is not in use for an extended period of time Always maintain the equipment clean for the next operation Do not modify any...

Страница 23: ...pment make sure that the liquid does not enter into the equipment To cleaning the equipment using the damp cloth with a mild soap The application of other solvent based cleaning products or sodium hyp...

Страница 24: ...e area to be cleaned leaving it to act for 5 minutes and drying it with a moist cloth check the instructions of use of the product with the manufacturer Note The registration at the Ministry of Health...

Страница 25: ...terilization process is described in the following sequence of illustrations This folding sequence isn t unintentional It makes its handling easier and it prevents contamination when the wrapping is o...

Страница 26: ...gh the sterilization process into the incubator in order to verify the incubator operation and also to control the viability of the spores used in the test check the BIs both the positive control and...

Страница 27: ...oclave s door can t be opened The autoclave isn t yet de pressurized Wait the autoclave depres surization Upon coming across any problem in operation follow the instructions below to check and repair...

Страница 28: ...28...

Страница 29: ...29...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ......

Отзывы: