5
Principles and fundamentals applied to the product functioning
Sterilization by autoclave consists of exposing the contaminated material to steam under high
pressures and temperatures, for the time needed. Humidity and heat irreversibly denature structural
enzymes and proteins, destroying microorganisms.
The process includes compression and decompression cycles as to provide easier contact between
the steam and the contaminated material. The pressure value ranges from 1.3 to 2.1 kgf/cm2 and
temperatures from 123°C to 133°C.
*
Optional
Description of Equipment
It’s an autoclave for sterilization by vapor, modern design, easy to operate, ideal for dentistry offices
(general practice, endodontia, pediatric dentistry, orthodontics, prosthesis, operative dentistry and
esthetics).
Its structure is made in carbon steel, with surface treatment through nanotechnology.
Smooth high gloss epoxy based paint polymerized at 250ºC with corrosion resistant treatment and
cleaning materials.
Its chamber is made of stainless steel
*
or aluminum
*
, offering long durability and resistance to the
constant variance of temperature, with capacity for 12 Liters.
It’s internally coated with heath isolating material, which optimizes its operation and reduces the
energy consumption.
It features an anti-vacuum valve, which prevents the door and safety seal from blocking, and it’s
automatically activated if an excessive pressure was detected. The door’s sealing system is made of
high-performance silicone, resistant to temperature and pressure changes, and its flexibility allows
the perfect sealing between surfaces in contact, preventing leakage.
The control panel contains LED and function buttons, graduated manometer for temperature and
pressure, ergonomic handle with automatic depressurization safety system, which prevents accidents
in case of unintentional opening.
Accompanied by a hose for outlet of steam, graduated cup, spare fuses and 2 trays made of stainless
steel or aluminum totally pierced to allow a good circulation of heat, ensuring excellent quality
in the sterilization.
Basic Principles of Autoclave’s Operation:
• Water supply for autoclave.
• Heating: Water steam followed by gradual increase of pressure.
• Sterilization: High pressures and temperatures through approximately 16 minutes.
• Depressurization: Steam release and relief of autoclave pressure.
• Drying: Material drying and cooling until achieving a temperature which allows handling.
IDENTIFICATION OF EQUIPMENT