24
Note
Changing in visual appearance inner the autoclave tank is a normal feature of the product
due to the heating of it.
Spots and oxidations signs may occur due to the continuous use and it is totally or partially
removed according to the recommended procedure as below:
In order to execute the internal cleaning of your equipment, sterilization chamber and
trays,we recommend using the product “SDS - 1000”. Product notified at ANVISA/MS.
Follow manufacturer’s directions for cleaning.
Product’s composition: inorganic-acids, fluoride and anionic tense-active based products.
Internal cleaning (chamber and stainless steel trays sterilization)
Warning
After the cleaning procedure perform a complete cycle of autoclaving and then eliminate
the water of the reservoir and dry the tub using a clean cloth.
Note
The registration at the Ministry of Health of “SDS - 1000” is performed separately from the
product described in this manual, as the “SDS - 1000” is not manufactured by Dabi Atlante.
Internal cleaning (aluminum sterilization chamber)
We recommend cleaning your autoclave monthly with a descaling product (HIGINDOOR 207 Reg.
MS: 310030079) which cleans the chamber. Apply a solution diluted at a 1:5 (1 part of the product
to 5 parts of water) of HIGINDOOR 207 in the area to be cleaned, leaving it to act for 5 minutes, and
drying it with a moist cloth. (check the instructions of use of the product with the manufacturer).
Note
The registration at the Ministry of Health of “HIGINDOOR 207” is performed separately
from the product described in this manual, as the “HIGINDOOR 207” is not manufactured
by Dabi Atlante.
Piping cleaning
Monthly the internal components of the autoclave must be cleaned and disinfected.
To clean the interior of your equipment, tubing, valves and filters, we recommend the use of “INOX
CLEAN” nº Reg. MS: 317130011.
Follow manufacturer’s directions for mixing the product.
Product’s composition: inorganic-acids, phosphate, acrilato de sódio. Mix the product following the
instructions given by the manufacturer.
Pour the solution in the chamber and start the sterilizing cycle. Repeat the sterilizing operation
using only distilled water. When these cycles are finished, remove the remnants of the cleaning
solution using a clean piece of cloth moistened with water.
Note
The registration at the Ministry of Health of “INOX CLEAN” is performed separately from the
product described in this manual, as the “INOX CLEAN” is not manufactured by Dabi Atlante.
Cleaning the filter
- The autoclave’s filter shall be cleaned daily.
- In order to clean the filter, remove it via the handle, and then wash it in running water.
- Don’t use sharp objects, which could damage the fabric.
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION