![D FASTER CytoFAST Top Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/d/faster-cytofast-top/faster-cytofast-top_operating-and-maintenance-manual_494325034.webp)
a D:GROUP Company
00055EN Rev.10 – 12/2011
33
10 TECHNICAL SPECIFICATIONS
Cabinetry
Steel-carbon steel varnished with epoxy powder by electro-spraying.
Perfectly air-tight structure with work area under negative pressure
Work chamber
Glazed AISI-304 stainless steel sheet with radiussed angles
Work surface
Perforated AISI 316L stainless steel sheet (removable)
Spillage tray
AISI 304 stainless steel, located under the work surface
Absolute filters (main/exhaust)
HEPA filters, tested with D.O.P. fumes; efficiency of over 99.999%
for particles of 0.3 micrometres. Under the work surface HEPA filter,
multidyhedral type, tested with D.O.P. fumes; efficiency over
99.99% for particles of 0.3 micrometre
Main motor-fan
Direct coupling centrifugal type. Designed to ensure a good
protective barrier even in the case of an exhaust motor-fan failure.
IP55 protection factor
Exhaust motor-fan
Direct coupling centrifugal type. Designed to keep the quantity of air
to be extracted constant and to create a protective barrier on the
front opening. Two speeds: a working speed and a double speed
which activates automatically in case of main motor-fan failure. This
second speed ensures the protection of the personnel and the
environment. IP32 protection factor
Speed regulation
Automatic regulation to keep constant air-flow speed (0.40 m/sec)
and the recycled air/extracted air ratio even in the presence of
progressive clogging of the three HEPA filters, up to pressure levels
nearing 36 mm of water column
Speed meter
Continuous monitoring and displaying of the laminar air-flow velocity
in the work chamber and of the protective barrier. These values are
measured by two fan-anemometers unaffected by variations in
temperature and humidity
Front glass
Safety glass, hinged to ensure ease of access for large items.
Equipped with gas spring(s) for opening/closing and holding the
window fully open (145° aperture)
Lighting
Fluorescent tubes situated outside the work area, medium lighting
system of the work surface > 800 lux
Control panel
Soft-touch keyboard on the front
Power supply
220-230V - 50 Hz single phas
Electrical system
according to EN-61010-1 + amendments
Testing
Electronic "light scattering" meter, fan anemometer, noise meter,
luxmeter, and smoke test at the factory.
Содержание FASTER CytoFAST Top
Страница 25: ...a D GROUP Company 00055EN Rev 10 12 2011 24 Picture 1 Picture 2 ...
Страница 38: ...a D GROUP Company 00055EN Rev 10 12 2011 37 13 A DIAGRAM FOR MAINTENANCE OPERATIONS Frontal view ...
Страница 39: ...a D GROUP Company 00055EN Rev 10 12 2011 38 Side view ...
Страница 40: ...a D GROUP Company 00055EN Rev 10 12 2011 39 13 B CytoFAST Top FRONTAL and SIDE DIAGRAM Frontal diagram ...
Страница 41: ...a D GROUP Company 00055EN Rev 10 12 2011 40 Side diagram ...
Страница 42: ...a D GROUP Company 00055EN Rev 10 12 2011 41 13 C SAFE WORKING AREA DIAGRAM ...