Post-fumigation cleaning
Having performed fumigation, the manipulation of the filters is safe: however, this manipulation should be
carried out by technicians wearing protective gloves and clothing.
Before using the cabinet again any possible remains of the foregoing procedure of vaporization and relevant
neutralization (which can show, for example, as white dust) must be removed from the work chamber.
ATTENTION: for cabinets that are not connected to the outside, but recycling the air in the
environment, make sure that ambient concentration does not exceed 2.5mg/m
3
, as
higher concentrations are harmful.
Faster S.r.l. refuses all responsibility for the performance of such sterilization procedures or of any
possible consequences deriving from it, reminding that the full responsibility for the operation lies
only and exclusively with the properly trained operator-user.
Содержание FASTER CytoFAST Top
Страница 25: ...a D GROUP Company 00055EN Rev 10 12 2011 24 Picture 1 Picture 2 ...
Страница 38: ...a D GROUP Company 00055EN Rev 10 12 2011 37 13 A DIAGRAM FOR MAINTENANCE OPERATIONS Frontal view ...
Страница 39: ...a D GROUP Company 00055EN Rev 10 12 2011 38 Side view ...
Страница 40: ...a D GROUP Company 00055EN Rev 10 12 2011 39 13 B CytoFAST Top FRONTAL and SIDE DIAGRAM Frontal diagram ...
Страница 41: ...a D GROUP Company 00055EN Rev 10 12 2011 40 Side diagram ...
Страница 42: ...a D GROUP Company 00055EN Rev 10 12 2011 41 13 C SAFE WORKING AREA DIAGRAM ...