PROBLEM
CAUSE
REMEDY
During work, an alarm sounds, the
laminar airflow works regularly,
and the velocity of the protective
barrier is reduced by about 30%.
Alarm: “barrier and exhaust fan"
The exhaust fan is out of order
See
the
above-mentioned
problem “Alarms: low barrier
velocity”.
IMPORTANT: if this problem
occurs, the material present in
the work area is not damaged,
the
laboratory
is sufficiently
protected
by
the
protection
barrier assured by the main fan
and by the design characteristics
of these cabinets.
If necessary, you can continue
working under protection, but
avoid quick movements and
large flames and do not place
large objects under the airflow.
During work, the laminar airflow
seems to work regularly, but the
alarm
“Barrier
velocity”
starts
sounding.
Exhaust and main filters are
clogged
Replace HEPA filters
Содержание FASTER CytoFAST Top
Страница 25: ...a D GROUP Company 00055EN Rev 10 12 2011 24 Picture 1 Picture 2 ...
Страница 38: ...a D GROUP Company 00055EN Rev 10 12 2011 37 13 A DIAGRAM FOR MAINTENANCE OPERATIONS Frontal view ...
Страница 39: ...a D GROUP Company 00055EN Rev 10 12 2011 38 Side view ...
Страница 40: ...a D GROUP Company 00055EN Rev 10 12 2011 39 13 B CytoFAST Top FRONTAL and SIDE DIAGRAM Frontal diagram ...
Страница 41: ...a D GROUP Company 00055EN Rev 10 12 2011 40 Side diagram ...
Страница 42: ...a D GROUP Company 00055EN Rev 10 12 2011 41 13 C SAFE WORKING AREA DIAGRAM ...