Spares/Options List

 

Q200 User Manual 

 

 

200120/Revision 10 

C-4

 

 

To clear this error, press the 

 key. The display will then show: 

 

Reset Parameters  No 

 
 
 
Pressing the 

 or 

 keys will scroll the response between yes and no. To clear the 

parameters (the recommended course of action) set the response to yes and press the 

 key; the following message will be displayed for two seconds: 

 

Parameters Cleared 

 
 
 
To leave the parameters as they are (not recommended) set the response to no and 
press the 

 key. 

 
C.2.4 I

NVALID 

D

ATE 

E

RROR

 

 
This error only occurs when setting the date. If an invalid date is entered 

e.g.

 31/02/94 

this message will be displayed for two seconds before returning to the date setting 
parameters. The display will show: 
 

Error: Invalid Date 

 
 
 
C.2.5 I

NVALID 

C

ALIBRATION 

E

RROR

 

 
If when performing a calibration, on pressing the 

 key to accept a calibration, the new 

calibration factor is less than 0.3 or greater than 3.0 this error message will be 
displayed for two seconds.  The display will show: 
 

Error: Invalid Cal 

 
 
 
Since an 'Invalid Cal' error leaves the calibration factor unchanged, it is possible to 
continue leak checking. However, entering the sample mode from standby by pressing 
the 

 key, the instrument will display a 'Warn: Not calibrated ' message to remind the 

user that the instrument is being used in a manner where the results obtained may not 
be reliable. 
 
'Invalid Cal' errors can result from poor instrument performance caused by 
contamination of the detector or sinter, a standard leak whose validity period has 
expired or the standard leak not having been turned on. 
 
C.2.6 N

L

EVELS 

S

ET 

E

RROR

 

 
This error only occurs when entering sample mode. If the user tries to enter sample 
mode without having set an alarm or a calibration level, this error message will be 
displayed for two seconds. The display will show: 
 
 

Error: No Levels Set 

 
 
This message is displayed for two seconds whereupon the display will then show: 
 
 
 

Содержание Q200

Страница 1: ...Uson Q200 User Manual Revison 10 Updated September 2018...

Страница 2: ...f such revisions or changes Copyright No part of this manual may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording for any purpose...

Страница 3: ......

Страница 4: ...2 03 1994 First Revision 3 N A 14 12 1994 Second Revision 4 1348 06 01 1995 Spares Updated 5 1415 25 01 1996 EC Revision 6 0051 02 07 1997 Sinter References 7 0263 08 11 1999 H M L Sinters Fixed expon...

Страница 5: ...OGRAMMING THE Q200 gives a detailed description of each program function and instructions for viewing and altering them Section 9 CALIBRATING THE Q200 describes the procedure for calibrating the Q200...

Страница 6: ...hanging the battery on the Q200 Appendix I Changing The Probe Sinter details the procedure for changing the Q200 s hand unit probe and sinter Appendix J Leak Rate and Concentration Formulae lists the...

Страница 7: ......

Страница 8: ...STALLATION REQUIREMENTS 4 1 4 1 REMOVING THE CASE LID 4 1 4 2 GAS SUPPLY 4 1 4 3 POWER SUPPLY 4 2 4 4 STORAGE AND OPERATING ENVIRONMENT 4 2 4 5 OPERATING MODES 4 2 4 5 1 PORTABLE INSTALLATION 4 2 4 5...

Страница 9: ...Q200 9 1 10 0 THE Q200 IN USE 10 1 10 1 STANDBY MODE 10 1 10 2 SAMPLE MODE 10 1 10 3 INSTRUMENT CALIBRATION 10 2 10 4 GENERAL TIPS ON LEAK DETECTION 10 3 10 5 ADJUSTING THE VOLUME 10 6 10 6 BACKGROUN...

Страница 10: ...ssories 3 1 Figure 2 Removing the Case Lid 4 1 Figure 3 Front Panel Layout 5 1 Figure 4 Hand Unit Display and Keyboard 6 1 Figure 5 Changing the Internal Gas Cylinder G 1 Figure 6 Changing the Battery...

Страница 11: ......

Страница 12: ...en of the instrument and all users are instructed in the use of the instrument and not to attempt access to the radioactive sources In the event of damage or loss occurring there may be a requirement...

Страница 13: ...to enter the detector Appendix D contains details of replacement seals that can be fitted to the sinter Fit the sinter into the front of the detector assembly Use a broad flat bladed screwdriver to ti...

Страница 14: ...t is not damaged Damage to the seal will compromise the performance of the instrument by allowing an excessive amount of air to enter the detector Appendix D contains details of replacement seals that...

Страница 15: ...ces to re enter a valid code If a valid code is not entered within three attempts the instrument will revert back to the Configure prompt 4 If the Select key is pressed with a valid code being display...

Страница 16: ...LY The instrument requires a supply of Argon that is at least 99 998 pure This is supplied with the instrument in a gas cylinder that conforms to BS 5045 3 B The gas cylinder should be filled to a pre...

Страница 17: ...ccident that damages the hand unit beyond repair or when the instrument has exceeded its useful life the unit must be disposed of through an approved contractor In the United Kingdom the manufacturer...

Страница 18: ...on Ltd Western Way Bury St Edmunds Suffolk United Kingdom IP33 3SP Transposed Harmonised Standards BS EN 61010 1 1993 Safety requirements for electrical equipment for measurement control and laborator...

Страница 19: ......

Страница 20: ...e detection principle uses a single Electron Capture Detector ECD utilising a low power radioactive source contained within the hand unit To measure for leaks the component under test is filled with a...

Страница 21: ......

Страница 22: ...00 are housed in the console case lid as illustrated in Figure 1 Figure 1 Q200 Case Lid Containing Accessories Gas cylinder spanner to allow the internal gas cylinder to be removed Battery charger and...

Страница 23: ...ery charger is located in the fuse drawer section of the charger mains socket inlet A range of standard leaks to be used during calibration and sinters to modify the operating range of the Q200 are av...

Страница 24: ...the case lid unlock the case release the buckles on the front of the case and raise the lid Whilst holding the case lid slide the lid to the left this will detach the lid from the base of the case 4 2...

Страница 25: ...moval and replacement of the battery is described in Appendix H The battery charger is designed for use on a flat surface adjacent to the instrument and with adequate clearance around the charger for...

Страница 26: ...ode of operation as the battery capacity will be slowly eroded Running the instrument in the detector clean mode while charging is also possible but a 24 hour charge is recommended in this case To rec...

Страница 27: ...has the effect of disabling the low gas warning See section 8 for details of how to program the instrument 4 5 2 2 RUNNING FROM THE EXTERNAL POWER SUPPLY The external power is supplied via the mains o...

Страница 28: ...00 Figure 3 Front Panel Layout 5 1 1 KEY SWITCH ORIENTATION The key switch has the following orientation Run key removed The user can make measurements select standby mode clean the detector zero the...

Страница 29: ...being displayed then that parameter will be reset to its default setting This key also cancels errors and exits functions such as the clean facility Clock Edit This key is used to either set the date...

Страница 30: ...unit consists of a rugged case This houses a sample probe detector with amplifier pump ten segment LED bar graph display three key switch panel and two LED indicators as illustrated in Figure 4 A serv...

Страница 31: ...eading taken since the last zero was performed When enabled the audio is still free running enabling the leak site to be located Pressing the key a second time will disable this function 6 2 HAND UNIT...

Страница 32: ...r calibration level with the alarm level taking precedence if both are set When the user is sampling and a leak is present the audio will build until the midpoint is reached then a two tone signal wil...

Страница 33: ......

Страница 34: ...the hand unit from its holder Turn on the gas supply by turning the gas bottle handle to the left or turn on the external gas supply Switch the instrument on via the on off rocker switch located in a...

Страница 35: ...d with a two second delay between each display update Warming Up Gas SF6 Units ml s Cal Off Al 1 0x10 6ml s Throughout the initialisation routine the hand unit warning LED is illuminated The hand unit...

Страница 36: ...Configure Diagnostics Set Date Time While in program mode the warning LED will be on to indicate the instrument is in program mode except when calibrating If the key is inserted in the key switch and...

Страница 37: ...ment may be calibrated are automatically changed to match the usable range of the instrument Appendix A contains a full list of the operating calibration and alarm ranges for all the inlet probes 8 1...

Страница 38: ...Configure menu use the keys and to exit the Configure menu press 8 2 1 CLEAN TIME The first parameter in the Configure menu is Clean Time 1 00 H M This parameter relates to the time for which the dete...

Страница 39: ...ibration level but in this case no two tone warning is given When results are printed the alarm level sets the criterion for printing Pass or Fail against individual results 8 2 3 TRACE GAS CONCENTRAT...

Страница 40: ...st Default Value Off 8 2 5 DISPLAY TYPE The fifth parameter in the Configure menu is Display Auto This parameter sets the manner in which leak rate information is presented on the digital display on t...

Страница 41: ...and keys this parameter can be toggled between no and yes To clear the program mode parameters select the Yes option and press the key on the hand unit The instrument will display Parameters Cleared T...

Страница 42: ...ue Internal 8 3 DIAGNOSTICS MENU The Diagnostics menu contains five parameters that are described in this section To enter the Diagnostics menu the user must press the key with the Diagnostics menu he...

Страница 43: ...in the Calibration menu then this parameter will display Recal Off 8 3 5 SOFTWARE VERSION NUMBER The fifth parameter in the Diagnostics menu is Version 200XXX 5 0 This parameter is the version number...

Страница 44: ...ds This message confirms that the Real Time Clock has been updated Date Set If on pressing the key the date entered is invalid e g 31 02 1999 an error message will be displayed for two seconds and the...

Страница 45: ...ngs press key to exit without changing any settings press the key When the user exits the Time sub menu a confirmation message will be displayed for two seconds This message confirms that the Real Tim...

Страница 46: ...Q200 Q200 User Manual 200120 Revision 10 8 11 8 5 EXITING THE MENU STRUCTURE Any of the menus mentioned in Section 11 can be exited by moving the key switch to or which will cause the instrument to en...

Страница 47: ......

Страница 48: ...Calibrate menu as described in section 8 1 3 The calibration routine is initiated by pressing the key on the console whilst in program mode Entering the program mode will cause the warning LED to be...

Страница 49: ...re reverting to its previous state before the key was pressed with the warning LED illuminated Valid Calibration If the calibration leak rate entered in the Calibrate menu varies greatly from the resp...

Страница 50: ...trument is powered up and the initialisation routine has been completed the display will show Standby The instrument should be put into standby mode if it is not going to be used for five minutes or m...

Страница 51: ...iate if the instrument is used in an essentially fixed installation a supply of clean air can be derived from a compressed air cylinder with a regulator or from a compressed air supply If a compressed...

Страница 52: ...entration of the gas in the vessel is representative of the bulk calibration gas by performing sequential calibrations and examining the calibration factor after each calibration and flushing the vess...

Страница 53: ...ed to stand for a specified period of time to allow the concentration of SF6 to build up in the space within the chamber surrounding the item under test The concentration of SF6 in this space is measu...

Страница 54: ...in the test chamber as a result of the incubation of a leaking test component Having zeroed the instrument in a clean background the instrument will display a true result because the SF6 concentration...

Страница 55: ...setting on the Q200 can be adjusted whilst in sample mode If the user presses either the or keys the new volume level will be displayed Volume Medium This message will be displayed for a two second p...

Страница 56: ...cific purpose To do this enter sample mode and clear the storage buffer by pressing and keys simultaneously If adding to any previous stored readings ignore this step To store a result whilst measurin...

Страница 57: ...8 66 67 Fail 4 33 33 Total 12 It should be noted that if no alarm level were set no pass fail results would be printed as shown below Q200 Results Gas SF6 Trace gas 100 Units ml s Al Off Cal 1 0x10 6m...

Страница 58: ...ided The pin outs for the RS232 connector are shown in Appendix E The printer should be set up to match the instrument s communication protocol 9600 baud 1 start bit 8 bit data 1 stop bit No parity 11...

Страница 59: ......

Страница 60: ...lay Direct Digital display of leak rate in ml s or ppm Bar graph display relative to alarm level or calibration level Alarm Audio and visual warning if the leak rate sampled exceeds the set alarm leve...

Страница 61: ...t The following tables indicate the leak rate and concentration measurement Calibration and Alarm ranges that are available for a range of specified Trace Gas concentrations Leak Rate Measuring Ranges...

Страница 62: ...inter L Sinter Range 0 0 x 10 8 1 8 x 10 5 0 0 x 10 6 1 8 x 10 3 0 0 x 10 5 1 8 x 10 2 0 0 x 10 4 1 8 x 10 1 Calibration Level 1 0 x 10 10 1 8 x 10 7 1 0 x 10 8 1 8 x 10 5 1 0 x 10 7 1 8 x 10 4 1 0 x...

Страница 63: ......

Страница 64: ...hould be left on charge overnight From time to time remove the battery and check for any cracks in the case and around the terminals Spare batteries can be obtained from your supplier see Appendix D f...

Страница 65: ...e the probe occasionally and check that it is not blocked Spare probes may be obtained from your supplier see Appendix D for more information B 6 CABLE Check the cable between the console and the hand...

Страница 66: ...uses Remove the fuse in the active position and replace with the spare fuse Refit the fuse drawer connect the charger to the mains supply and re check If the unit still does not perform correctly seek...

Страница 67: ......

Страница 68: ...t C 1 1 EPROM FAULT This fault will only occur at power up The display will show Fault EPROM If this fault occurs consult your supplier C 1 2 MEMORY FAULT This fault will only occur at power up The di...

Страница 69: ...0 C in normal sample mode or exceeds 120 C in clean mode the heater control circuit could be faulty The display will show Fault Heater If this fault occurs consult your supplier C 1 8 PRT OPEN CIRCUIT...

Страница 70: ...up if any stored program parameters have been corrupted The display will show Error Program To clear this error press the key The display will then show Reset Parameters No Pressing the or keys will...

Страница 71: ...alibration factor is less than 0 3 or greater than 3 0 this error message will be displayed for two seconds The display will show Error Invalid Cal Since an Invalid Cal error leaves the calibration fa...

Страница 72: ...an always be cancelled If as a result of the warning the key switch is set to Program and a calibration is attempted and no calibration level has been set the display will show No cal leakrate set Thi...

Страница 73: ...e Diagnostics menu If this does not improve the detector frequency the inlet either sinter or capillary should be changed Appendix D contains details of the part numbers of the seals and inlets that m...

Страница 74: ...NG This warning can only occur in sample mode when the user tries to store more than 100 readings The display will show Warn Storage Buffer An intermittent audio warning will sound and the warning LED...

Страница 75: ......

Страница 76: ...inter O ring Seal 2 674 Capillary Probe Kit Range 10 10 10 7 100 Trace Gas 200128 Capillary Probe Seals O Ring Needle Probe Disc Seal 2 671 37107 40 Column printer with RS232 printer cable 200123 RS23...

Страница 77: ...Q200 User Manual Spares Options List 200120 Revision 10 D 2 Gas bottle filling jig American 99030 Standard leak kit Leak rate to be specified 10050 Inline battery fuse T2A 250V 1 3 291...

Страница 78: ...and the analogue output for connection to a printer and chart recorder respectively Pin Signal name function Direction of signal 1 2 RxD Not used 3 Txd Transmitted data From 200 to printer 4 5 6 DSR 2...

Страница 79: ......

Страница 80: ...leaks in high voltage electrical switchgear Fume cupboard and glove box checks Location of leaks in military components e g missiles mines and torpedoes Leak testing fire extinguishers Leak testing of...

Страница 81: ......

Страница 82: ...le lid accessible when the console lid is raised Use gas cylinder spanner provided to unscrew cylinder connection Release the cylinder retaining straps and withdraw cylinder Fit new cylinder ensuring...

Страница 83: ......

Страница 84: ...attery is disconnected to reconnect the battery the procedure is as follows Turn off power Unlock the turn fasteners on the console cover 3 positions raise console cover and engage support arm Unlock...

Страница 85: ...e the connector that runs into the main cable Lift battery assembly out of case Install new battery assembly in the console Locate the battery lead and connect it to its mating connector lead that is...

Страница 86: ...d unit nozzle illustrated in Figure 8 To change the instrument s sinter the procedure is as follows Turn off the power and gas supplies and allow the instrument to cool Unscrew the hand unit nozzle Re...

Страница 87: ...he sinter into the detector assembly using a broad bladed screwdriver to ensure that the slot in the front of the sinter in not damaged Replace the nozzle Figure 8 Location of Sinter Warning The detec...

Страница 88: ...g formulae J 1 LEAK RATE IN GRAMS PER YEAR GM YEAR gm year leak rate in ml per second gas molecular weight 31536000 24500 where gas molecular weight 146 for SF6 31536000 Number of seconds in a year 24...

Страница 89: ......

Страница 90: ...Standby Sam ple Calibrate CleanTime Configure Diagnostics SetDate Time Date Time Hours M inutes Day Month Battery Tem perature Frequency Recal Version Reset Parameters Al TraceGas AutoZero Display Ga...

Страница 91: ......

Страница 92: ...n factor that is applied to all future readings Cal Factor The calibration factor is the ratio by which the actual measurement of leak rate obtained from a standard leak differs from notional measurem...

Страница 93: ...ered since last time the key was pressed With peak hold selected the audio level still indicates the instantaneous signal level Recal Relates to the time remaining before re calibration is due Recal I...

Страница 94: ...2 C 6 C 7 C 8 clock 5 2 8 9 8 10 concentration iii 8 4 10 1 A 2 J 1 ii contamination 8 3 8 5 10 1 10 7 B 1 ii cylinder 1 5 2 1 3 1 4 3 4 4 A 1 B 2 C 2 G 1 D date 5 2 8 1 8 9 8 10 B 2 C 4 detector1 1 2...

Страница 95: ...robe iii 1 6 6 1 9 1 9 2 10 6 10 7 B 2 I 1 ii R re calibration 8 2 8 8 B 1 C 8 ii re calibration interval 8 2 8 8 B 1 C 8 reset parameters 6 2 S sample 5 2 6 1 6 2 8 3 9 1 10 1 10 6 10 7 C 2 C 5 C 6 C...

Отзывы: